DOI QR코드

DOI QR Code

Development of multi-colorimeter module for low-cost urinalysis strip readers

저가형 요분석 시스템의 다중 광 검출 모듈개발

  • Ye, Soo-Young (BK21 Medical Science Education Center, School of Medicine, Pusan National University) ;
  • Jeon, Yong-Uk (Dept. of Interdisciplinary Program in Biomedical Engineering, Graudated School Pusan National University) ;
  • Jeong, Do-Un (Division of Computer & Information Engineering, Dongseo University) ;
  • Jeon, Gye-Rok (Dept. of Biomedical Engineering, School of medicime, Pusan National Universiry) ;
  • Ro, Jung-Hoon (Dept. of Biomedical Engineering, School of medicime, Pusan National Universiry)
  • 예수영 (부산대학교 의학전문대학원 BK21 고급의료인력양성사업단) ;
  • 전용욱 (부산대학교 의학대학 의공학협동과정) ;
  • 정도운 (동서대학교 컴퓨터정보공학부) ;
  • 전계록 (부산대학교 의학전문대학원 의공학교실) ;
  • 노정훈 (부산대학교 의학전문대학원 의공학교실)
  • Published : 2008.09.30

Abstract

An optic module system is developed adopting multiple colorimetry units for the measurement of multi-pad urinalysis dipsticks. Multiple photometry system instead of moving mechanisms has the advantages of system reliability and simplicity as well as economic aspects due to the recent development of economic color light emitting diodes and stable photo sensors. An integration amplifier with programmable integration time, a current source circuit with selectable and stable current settings were connected through analog multiplexers to thirty light emitting diodes for illumination and ten photo transistors for reading each strip pad. All the circuits are controlled by a microprocessor through a simple set of serial communication commands. The detect ability is eighteen times better than the minimum color difference of the test grading which is 0.013 in urobilinogen in the color space defined in this paper.

Keywords

References

  1. L.E. Armstrong, C.M. Maresh, J.W. Castellani, M.F. Bereron, R.W. Kenefick, K.E. LaGassee and D. Riebe, "Urinary indices of hydration status.", Int. J. Sport Nutr., vol. 4, pp. 265-279, 1994 https://doi.org/10.1123/ijsn.4.3.265
  2. L.E. Armstrong, J.A. Soto, F.T. Hacker, D.J. Casa, S.A. Kavouras and C.M. Maresh, "Urinary indices during dehydration, exercise, and rehydration", Int. J. Sport Nutr., vol. 8, pp. 345-355, 1998 https://doi.org/10.1123/ijsn.8.4.345
  3. William E. Neeley, "eflectance photometer for multilayer dry film slides", Anal. Chem., vol. 56, pp. 742-745, 1984 https://doi.org/10.1021/ac00268a033
  4. Adam Zipp and W. E. Hornby, "Solid-phase chemistry : Its principles and applications in clinical analysis", Talanta, vol. 31, no. 10B, pp. 863-877, 1984 https://doi.org/10.1016/0039-9140(84)80215-5
  5. 유동주, 정태화, 민홍기, "건습시약 스트립을 사용한 임상화학분석장치 개발[I]", 의공학회지, 제13권, 제 4호, pp. 229-305, 1992
  6. John A. Lott and Elias Khabbaza, "Haemoglobin analysis on whole blood by reflectance photomerty", J. of Automat. Chem., vol. 7, no. 4, pp. 197-200, 1985 https://doi.org/10.1155/S1463924685000438
  7. 승진, "색도 좌표 변환 및 분류 기법을 이용한 휴 대형 요분석 시스템의 구현", 부산대학교 박사학위 논문, 2002
  8. Roy S. Berns, "Principles of color technology", John Wiley & Sons, Inc., pp. 222-226, 2000

Cited by

  1. Study on a novel core module based on optical fiber bundles for urine dry-chemistry analysis vol.88, pp.9, 2017, https://doi.org/10.1063/1.4990686