Case Study of Quality Assurance for MDCT Image Quality Evaluation Using AAPM CT Performance Phantom

AAPM CT 성능 평가용 팬텀을 이용한 전산화단층촬영의 영상 평가를 위한 정도관리 사례 연구

  • 장근조 (예수병원 영상의학과) ;
  • 권대철 (서울대학교병원 영상의학과)
  • Published : 2007.07.28

Abstract

The increasing use of computed tomography (CT) as a diagnostic tool creates the need from and efficient means of evaluating the performance of the CT scanner now in use. Accordingly, acceptance testing and quality assurance of CT is of great importance. The aim of this study is to analyze of AAPM CT performance phantom in the CT accreditation program. The modular phantom offers the CT system with which to measure eight performance parameters. The parameters are listed of CT attenuation coefficient of water, noise, uniformity, spatial resolution, contrast resolution, slice thickness (5 and 10 mm), artifacts and alignment. The phantom evaluation was done by two radiologists. The acceptance testing protocol described here in demonstrates the successful of the guidelines for the quality assurance using AAPM CT performance phantom. We need to be upgraded for the CT image quality and make the standard reference of the quality assurance in the CT.

전산화단층촬영(CT)은 영상의학에서 매우 유용한 검사의 진단법으로 적절한 정도관리에 의한 영상의 평가가 필요하다. CT의 정도관리 항목에서 영상의 질을 결정하는 중요한 요소를 위해 AAPM CT 팬텀으로 영상검사를 수행하였다. 정도관리 평가항목은 "특수의료장비의 설치 및 운영에 관한 규칙"에서 정한 기준으로 물의 CT 감약계수, 노이즈, 균일도, 공간분해능, 대조도 분해능, 절편 두께, 인공물 존재 유무를 평가하였다. 획득한 영상은 규칙의 합격기준에 충족하였다. CT 영상의 질을 최적으로 유지하기 위해서는 지속적으로 팬텀 및 임상검사를 통해 영상의 질을 평가하기 위해 정도관리를 시행하여야 한다.