The Evaluation of Image Quality and Radiation Dose in Multi-Detector CT
MDCT에서 화질과 방사선량에 관한 연구
- Han, Dong-Kyoon (Dept. of Radiological Science, College of Health Science, Eulji University) ;
- Yang, Han-Joon (Dept. of Radiological Science, College of Health Science, Eulji University) ;
- Kim, Moon-Chan (Dept. of Diagnostic Radiology, Samsung Seoul Hospital) ;
- Ko, Shin-Gwan (Dept. of Radiological Science, College of Health Science, Eulji University)
- Published : 2007.06.30
Abstract
The Purpose of this study is to suggest the basic data for making good quality image and maintaining equipment homeostasis by accepting image quality evaluation and radiation dose evaluation in Multi-detector CT. In this study we surveyed 14 CT equipments in Seoul. The results obtained were as follows ; CT number was
본 연구는 MDCT장치에 있어 화질평가와 선량평가 방법을 통하여 장치의 항상성을 유지하며 양질의 의료영상을 제공하는 기초자료로 사용함과 더불어, MDCT의 각 장비에 따른 선량을 평가하고 제시하고자 서울소재 14대의 MDCT장치를 대상으로 시행하여 그 기준을 정량화 하였다. MDCT를 이용한 화질측정 즉, CT number와 노이즈, 균일도, 공간분해능, 대조도 분해능과 선량 측정 즉, CTDI와 CTDIw, CTDIw/100 mAs의 결과는 다음과 같았다. CT number는 평균