Effectiveness of Bellyboard Device for Displacement of Small Bowel in Pelvic Irradiation

골반 방사선치료 시 소장의 위치변화를 위한 벨리보드의 유용성

  • Lee, Rena (Department of Radiation Oncology, School of Medicine, Ewha Womans University) ;
  • Lee, Kyung-Ja (Department of Radiation Oncology, School of Medicine, Ewha Womans University) ;
  • Suh, Hyunsuk (Department of Radiation Oncology, School of Medicine, Ewha Womans University)
  • 이레나 (이화여자대학교 의학전문대학원 방사선종양학교실) ;
  • 이경자 (이화여자대학교 의학전문대학원 방사선종양학교실) ;
  • 서현숙 (이화여자대학교 의학전문대학원 방사선종양학교실)
  • Received : 2007.08.13
  • Accepted : 2007.11.12
  • Published : 2007.12.31

Abstract

Various techniques were evaluated to determine the best method for reducing small bowel involvement in pelvic irradiation. Fourteen patients receiving radiation in pelvic area were enrolled for this study. Five sets of small bowel images were obtained. Patients were positioned on a simulation couch with full bladder in prone and supine positions and 2 sets of images were taken. Then they were asked to empty their bladder and 2 sets of images were taken in prone and supine positions. A belly board device (BBD) was placed and one set of images was obtained. Using a software, the area of small bowel inside treatment field was contoured, measured, and analyzed. In both full and empty bladder cases, small bowel area reduction was observed in prone position as compared to supine position. Especially statistically significant reduction is noted in lateral film. An average decreases of 13% in PA and 26% in lateral direction were noted with bladder distention as compared to empty bladder. With the use of BBD for empty bladder, a significant reduction of $62.8{\pm}27.1%$ and $63.1{\pm}32.9%$ in PA and lateral directions were observed as compared to without BBD in prone position, respectively. In conclusion, the best sparing of small bowel concerning the area included in the treatment fields was achieved with BBD in prone position with empty bladder. However, further reduction is expected if the bladder was filled fully because the analysed data with empty vs full bladder study shows increased sparing of small bowel with distended bladder.

골반 방사선치료 시 방사선조사부위에서 소장을 제외하여 방사선에 의한 소장의 독성 감소에 가장 효율적인 방법을 알아보기 위하여 본 연구가 시행되었다. 연구대상은 자궁경부암, 자궁내막암, 직장암, S-결장암과 난소암으로 수술 후 골반에 방사선치료를 받은 환자 14명이다. 환자는 모의치료를 시행하기 1시간 30분 전에 150 ml의 바륨과 방광을 채우기 위해 물 500 ml를 마시고 환자의 자세변화와 벨리보드 사용 유무에 따라 5종류의 조영촬영을 하였다. 첫 번째는 방광을 채운 후 누운 자세로 전 후면과 측면을 촬영한다. 두 번째는 방광을 채운 상태에서 엎드린 자세로, 세 번째는 방광을 비운 후 누운 자세로, 네 번째는 방광을 비운 후 엎드린 자세로, 다섯 번째는 방광을 비운 후 벨리보드를 사용하여 엎드린 자세로 각각의 전 후면과 측면을 촬영한다. 모든 조영촬영 사진에서 방사선조사부위에 포함되는 소장의 용적을 측정하였다. 결과는 방광을 채우거나 비운상태 모두 엎드린 자세가 누운 자세보다 소장의 용적이 감소되었으며 특히 측면사진에서 더욱 감소되었다. 엎드린 자세에서는 방광을 채운 상태의 소장 용적이 방광을 비운상태에 비해 전후면 사진에서 13%, 측면 사진에서 26%가 감소되었다. 방광을 비운 후 벨리보드를 사용한 상태의 소장 용적이 벨리보드를 사용하지 않고 엎드린 자세에 비해 전 후면에서 $62.8{\pm}27.1%$, 측면에서 $63{\pm}32.9%$가 감소되었다. 결론적으로 골반에 방사선치료를 시행할 경우 방광을 비운 후 벨리보드를 사용함으로써 소장의 용적을 최소화시킬 수 있었다.

Keywords