The Acceptance Testing of 5 Mega Pixels Primary Electronic Display Devices and the Study of Quality Control Guideline Suitable for Domestic Circumstance

5 Mega 화소 진단용 전자표시장치 인수검사 및 국내 실정에 적합한 정도관리 가이드라인 연구

  • Jung, Hai-Jo (Research Institute of Radiological Science, Yonsei University) ;
  • Kim, Hee-Joung (Department of Radiological Soience/Radiology, Yonsei University College of Health Science/Medicine) ;
  • Kim, Sung-Kyu (Department of Therapeutic Radiology & Oncology, College of Medicine, Yeungnam University)
  • 정해조 (연세대학교 방사선의과학연구소) ;
  • 김희중 (연세대학교 보건과학대학 방사선학과) ;
  • 김성규 (영남대학교 의과대학 방사선종양학교실)
  • Published : 2007.06.30

Abstract

In June 2005, Yonsei University Medical Center, Severance Hospital upgraded a full-PACS system by adding twenty (5 mega pixels) Totoku ME511L flat panel LCD display devices for diagnostic interpretation purposes. Here we report upon the quantitative (or visual) acceptance testing of the twenty Totoku ME511L display devices for reflection, luminance response, luminance spatial dependency, resolution, noise, veiling glare, and display chromaticity based on AAPM TG 18 report. The tools used in the tests included a telescopic photometer, which was used as a colorimeter, illuminance meter, light sources for reflection assessment, light-blocking devices, and digital TG18 test patterns. For selected 8 flat panel displays, mean diffuse reflection coefficient ($R_d$) was $0.019{\pm}0.02sr^{-1}$. In the luminance response test, luminance ratio (LR), maximum luminance difference ($L_{max}$), and deviation of contrast response were $550{\pm}100,\;2.0{\pm}1.9%\;and\;5.8{\pm}1.8%$, respectively. In the luminance uniformity test, maximum luminance deviation was $14.3{\pm}5.5%$ for the 10% luminance of the TG18-UNL10 test pattern. In the resolution test with luminance measurement method, percent luminance (${\Dalta}L$) at the center was $0.94{\pm}0.64%$. In all cases of noise testing, rectangular target In every square in the three quadrants was visible and all 15 targets except the smallest one in the every corner pattern and the center pattern. The glare ratio (GR) was $12,346{\pm}1,995$. The color uniformity, (u',v'), was $0.0025{\pm}0.0008$. Also, the research results of qualify control guideline of primary disply devices suitable for domestic circumstance are presented All test results are in-line with the criteria recommended by AAPM TG18 report and are thus fully acceptable for diagnostic image interpretation. As a result, the acceptance testing schedule described provides not only an acceptance standard but also guidelines for quality control, optimized viewing conditions, and a means for determining the upgrading time of LCD display devices for diagnostic interpretation.

2005년 6월 연세의료원 세브란스병원에서는 진단목적으로 20대의 5 mesa 화소 Totoku ME511L 평판 LCD 표시장치를 추가하여 PACS를 확장하였다. 20대의 Totoku ME511L 표시장치를 대상으로 반사, 휘도 반응, 휘도 공간 의존, 분해능, 잡음, 베일링 그레어, 그리고 색도 항목에 대하여 AAPM TG 18보고서에 기준하여 정량적인(또는 시각적인) 인수검사를 실시하였다. 측정에 사용된 장치는 색도를 측정할 수 있는 만원경식 휘도계, 조도계, 반사 측정에 필요한 빛 선원, 빛 차단 장치, 그리고 TG18 테스트 패턴들이었다. 선택된 8대의 표시장치에 대한 평균 확산 반사계수($R_d$)는 $0.019{\pm}0.02sr^{-1}$이었다. 휘도 반응 검사에서, 휘도비(LR), 최대 휘도 변이($L_{max}$), 그리고 최대 대조도 반응 변이는 각각 $550{\pm}100,\;2.0{\pm}1.9%$, 그리고 $5.8{\pm}1.8%$이었다. 휘도 균일도 검사에서, 최대 휘도 변이는 TG18-UNL10 테스트 패턴의 10% 휘도에 대하여 $14.3{\pm}5.5%$이었다. 분해능 측정은 휘도 측정법으로 수행하였으며 중심에서 퍼센트 휘도(${\Dalta}L$)는 $0.94{\pm}0.64%$이었다. 시각적인 방법의 잡음측정의 모든 경우에서, 중앙 및 주위 4사분면에서 가장 작은 것을 제외하고 15개의 모든 정사각형의 표적을 인식할 수 있었다. 글레어 비(GR)은 $12,346{\pm}1,995$이었다. 최대 색도 균일도 지표(D)는 $0.0025{\pm}0.0008$이었다. 또한 국내 실정에 적합한 진단용 표시장치의 정도관리 가이드라인 연구 결과를 제시하였다. 모든 인수검사 결과는 AAPM TG18 보고서 기준에 포함되어 진단 목적으로 사용하기에 적합하였다. 결론으로 인수검사는 표시장치가 진단에 사용하기에 적합한 성능임을 알려줄 뿐만 아니라 정도관리의 가이드라인, 최적 판독 환경, 그리고 표시장치의 교체시기를 결정하여주는 중요한 역할을 할 수 있다.

Keywords