Development of 3 Channel Biomedical Signal Measurement System for Mac-yule

맥율용 3채널 생체신호 계측시스템 개발

  • Byeon, M.K. (Dept. of Electronics, Myongji University) ;
  • Kim, H.J. (Dept. of Electronics, Myongji University) ;
  • Jang, J.K. (Dept. of Electronics, Myongji University) ;
  • Han, S.W. (Dept. of Electronics, Myongji University) ;
  • Huh, W. (Dept. of Electronics, Myongji University)
  • 변미경 (명지대학교 전자공학과) ;
  • 김현준 (명지대학교 전자공학과) ;
  • 장준근 (명지대학교 전자공학과) ;
  • 한상휘 (명지대학교 전자공학과) ;
  • 허웅 (명지대학교 전자공학과)
  • Published : 2007.03.31

Abstract

In this paper, we developed a Mac-Yule measurement system which consider psychological stable state of patience. The developed system consist with a hardware device that can derive a EEG, respiration and pulse wave, and a software which acquire a biological signal and signal processing The EEGs are derived with bipolar method from frontal head. The respiration signals obtain from nasal front with a transducer which consist with thermistor bridge. The pulse waves are detected from earlobe with photoplethysmograph method. A power spectrum of EEG are used as the decision parameters of psychological stable state of patience. The decision of Mac-Yule are defined as origin text method that of numbers of pulse to 1 respiration period. As the results of experiment with developed system, we could have a spectrum band discretion of EEG signal, stable respiration signal detection and automatic gain controlled pulse signal with realtime. And then, we could detect Mac-Yules from processed signals.

본 연구에서 심리적으로 안정된 상태에서 맥율을 측정할 수 있는 장치를 개발하였다. 개발된 시스템은 뇌파, 호흡파, 맥동파를 검출하는 하드웨어장치와 이들 신호를 획득하고 처리하는 소프트웨어로 구성하였다. 뇌파는 전두부에서 쌍극형으로 유도하였고, 호흡은 서미스터 브리지를 이용하여 구성된 변환기를 사용하여 비강 전부에서 유도하였으며 맥동파는 귀볼에서 유도한 용적맥파를 사용하였다. 피검자의 심리적 안정된 상태의 판정은 뇌파의 스펙트럼을 이용하였다. 맥율의 결정은 원전에 따라 1호흡 당 맥동수를 사용하였다. 개발된 장치를 사용하여 맥율검출 실험을 한 결과, 뇌파의 주파수 대역별 구분, 안정된 호흡신호의 검출과 이득 조절이 되는 용적맥파의 검출이 실시간으로 가능하였다. 그리고 검출된 신호로부터 맥율을 검출할 수 있었다.

Keywords