The Comparison of the Accuracy upon the Manual Torque Generating Devices for the Implant

임플란트에 사용되는 수동 토크발생장치의 정확성 비교

  • Kim, Dong-Hwan (Department of Prosthodontics, College of Dentistry, Chosun University) ;
  • Kim, Han-Su (Department of Prosthodontics, College of Dentistry, Chosun University) ;
  • Oh, Sang-Ho (Department of Prosthodontics, College of Dentistry, Chosun University) ;
  • Chung, Chae-Heon (Department of Prosthodontics, College of Dentistry, Chosun University)
  • 김동환 (조선대학교 치과대학 보철학교실) ;
  • 김한수 (조선대학교 치과대학 보철학교실) ;
  • 오상호 (조선대학교 치과대학 보철학교실) ;
  • 정재헌 (조선대학교 치과대학 보철학교실)
  • Published : 2007.09.30

Abstract

Various methods and torque generating devices are developed and employed to apply proper torque clinically. In this study, clinically used torque control devices are measured whether uniformed and precise torque are generated. Digital torque gauge are employed for measurement and torque driver, torque limiting device, torque indicating device are used for torque generating devices. ${\chi}^2$ distribution was formed and One-way ANOVA(Turkey test) was performed to measure torque values generated by each torque generating device. In the results, all companies have been shown slight errors and deviations. This indicates a difficulty of applying precise torque. Therefore, it would be recommended torque generating device should be checked whether uniformed and precise torque can be generated and an error should be corrected.

Keywords

References

  1. Taylor A. Prosthodontic problems and limitations associated with osseointegration. J Prosthet Dent 1998;79:74-78 https://doi.org/10.1016/S0022-3913(98)70197-0
  2. Jemt T, Pettersson P. A 3-year follow-up study on single implant treatment. J Dent 1993;21:203-208 https://doi.org/10.1016/0300-5712(93)90127-C
  3. McGlumphy EA. Kerby RE, Elfers CL. A comparison of screw preload for the single tooth implant. J Dent Res 1994;73:203-209
  4. Goodacre CJ, Bernal G, Rugncharassaeng K, Kan JY. Clinical complications with implants and implant prostheses. J Prosthet Dent 2003;90:121-132 https://doi.org/10.1016/S0022-3913(03)00212-9
  5. Martin WC, Woody RD, Miller AW. Implant screw rotations and preloads for four different screw materials and surface. J Prosthet Dent 2001;86(1):24-32 https://doi.org/10.1067/mpr.2001.116230
  6. Binon PP, Sutter F, Beaty K, Brunski J, Gulbransen H, Weiner R. The role of screws in implant systems. Int J Oral Maxillofac Implants 1994;9:8-63
  7. Miller RB, McGlumphy EA, Kerby RE. Comparison of abutment screw preload in different implant designs. J Dent Res 1994;73:202-208
  8. Jomeus L, Jemt T, Carlsson L. Loads and designs of screw joints single crowns supported by osseointegrated implants. Int J Oral Maxillofac Implants 1992;7:353-359
  9. Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac implants 1995;10:529-536
  10. Burguette RL, John RB, King T. Patterson EA. Tightening characteristic for screwed joint in osseointegrated dental implant. J Prosthet Dent 1994;71:592-600 https://doi.org/10.1016/0022-3913(94)90443-X
  11. Motosh N. Development of design charts for bolts preloaded up to the plastic range. J Eng Ind 1976;98:849-851 https://doi.org/10.1115/1.3439041
  12. Winkler S, Ring K, Ring JD, Boberick KG. Implant screw mechanics and the settling effect. J Oral Implantol 2003;29:242-245 https://doi.org/10.1563/1548-1336(2003)029<0242:ISMATS>2.3.CO;2
  13. Binon P, Sutter F, Beaty K. Brunski J. Gulbransen H. Weiner R. The role of screws in implant systems. Int J Oral Maxillofac Implants 1994;9:48-63
  14. Tan KB, Nicholls JI. Implant prostheses screw joint compressive preload with hand torque driver vs. electri torque driver. J Dent Res 1994;73:203-209
  15. McGlumphy EA, Mendel DA, Holloway JA. Implant screw mechanics. Dent Clin North Am 1998;42:71-89
  16. Dellinges M, Curtis D. Effects of infection control procedures on the accuracy a of mechanical torque wrench system for implant restorations. J Prosthet Dent 1996;75:93-98 https://doi.org/10.1016/S0022-3913(96)90424-2
  17. Rogers D. Engineering a device for instaIling dental implants. Mechanical engineering 1992;114(4):72-73
  18. Standlee JP, Caputo AA. Accuracy of electric torque limiting device for implants. Int J Oral Maxillofac Implants 1999;14:278-281
  19. Ricardo Mitrani, Jack I, Keith M, Tsun Ma. Accuracy of electronic implant torque controllers following time in clinical service. Int J Oral Maxillofac Implants 2001;16:394-399
  20. Jont P, Angelo A, Ming-ya J. Tao T. Accuracy of mechanical torque limiting devices for implants. Int J Oral Maxillofac Implants 2002;17:220-224
  21. Jaarda MJ, Razzoog ME, Gratton DG. Effect of preload torque on the ultimate tensile strength of implant prosthetic retaining screws. Implant Dent 1994;3:17-21 https://doi.org/10.1097/00008505-199404000-00002
  22. Goheen KL,Vermilyea SG, Vossoughi J, Agar JR. Torque generated by handle screw drivers and mechanical torquing devices for osseointegrated implants. Int J Oral Maxillofac Implants 1994;9:149-155
  23. Binon PP. Evaluation of the effectiveness of a technique to prevent screw loosening. J Prosthet Dent. 1998;79:430-432 https://doi.org/10.1016/S0022-3913(98)70157-X
  24. Dellinges MA, Tebrock OC. A measurement of torque values obtained with hand-held drivers in a simulated clinical setting. Int J Prosthodont 1993;2:212-214 https://doi.org/10.1111/j.1532-849X.1993.tb00411.x
  25. Sherwood RL Jr, Sullivan DY. Concepts and techniques of single-tooth implant restorations. Esthet Dent Update 1991;2:16-22
  26. Gutierrez J. Nicholls JI, Libman WJ, Buston TJ. Accuracy of the Implant Torque Wrenches Following Time in Clinical Service. Int J Prosthodont 1997;17:562-567