Abstract

The MM22 microtron has used as a cancer therapy machine from Nov. 1986 to Feb. 2006. This machine was moved and installed to a radiation research center to use as an education and research tool from treatment machine because of aging of MM22 microtron. In this paper, for extracting the electron beam from microtron, operation principle of the microtron, system characteristics of each module, and pulse structures were reviewed. The beam extraction and measurement were performed after measuring pulses of each major module and extraction trials in the beam line. After finishing the movement of MM22 microtron, the 30mA target current in the case of 10 MV X-ray beam was extracted and the beam flatness of radiation distribution was acquired within 3% error ratio after 100 MU was irradiated on X-omatV Film at SSD 100 cm and field size $10{\times}10cm^2$. As a result, the microtron movement and new installation was performed with success.

MM22 마이크로트론은 1986년도 11월부터 2006년 2월까지 암 환자를 위한 방사선 치료 장비로 사용되었다. 장비의 노후로 치료 장비에서 연구 및 교육용으로 전환하기 위해 방사선의학연구센터로 이전 설치하였다. 본 논문에서는 이전 설치 한 후 빔 인출을 수행하기 위해 마이크로트론의 동작원리, 시스템을 구성하는 각 장치의 특성을 분석하여 보았고 주요 부분의 파라메타인 펄스구조의 특징을 살펴보았다. 실제, 각 주요 시스템의 펄스를 측정하였고 빔 인출부, 빔 라인 및 최종단인 타겟에서 빔 인출 기법을 통하여 빔 인출 및 빔 측정을 수행하였다. 이전 설치 후 10 MV X-선의 경우 최종 단 치료기에서 30 mA 타겟 전류를 인출하였고, 필름을 SSD 100 cm, $10{\times}10cm^2$ 조사면에 놓고 100 MU 방사선을 조사하였다. 조사면의 방사선분포의 평탄도 측정 결과 3%이내로 안정적인 빔을 인출하여 이전설치가 성공적으로 수행되었음을 확인할 수 있었다.

Keywords