A Study of Parallel Test Among Three ADVIA 2120 System
3대의 ADVIA 2120 System 평행시험에 대한 연구
- Chang, Sang-Wu (Tech Infor & QA Div, Green Cross Reference Laboratories) ;
- Cho, Eun-Hae (Department of Clinical Hematology Laboratory, Green Cross Reference Laboratories) ;
- Kim, Nam-Yong (Tech Infor & QA Div, Green Cross Reference Laboratories) ;
- Chu, Kyung-Bok (Department of Clinical Hematology Laboratory, Green Cross Reference Laboratories) ;
- Lee, Suk-Jong (Department of Clinical Hematology Laboratory, Green Cross Reference Laboratories) ;
- Hong, Sung-No (Department of Medical Laboratory Science, Dongnam Health College) ;
- Oh, Jong-Do (Department of Laboratory Medicine, Gyungbug University Hospital)
- 장상우 (녹십자의료재단, 고객지원 및 QA부) ;
- 조은해 (녹십자의료재단, 혈액검사실) ;
- 김남용 (녹십자의료재단, 고객지원 및 QA부) ;
- 추경복 (녹십자의료재단, 혈액검사실) ;
- 이숙정 (녹십자의료재단, 혈액검사실) ;
- 홍성노 (동남보건대학, 임상병리과) ;
- 오종도 (경북대학병원 진단검사의학과)
- Published : 2006.04.30
Abstract
Parallel testing means ordering a number of tests at the same time so abnormalities in any of the tests can be found quickly and used in making the diagnosis. This is a good medical strategy to eliminate diseases and it is relatively inexpensive if all the tests are potential sources of information and performed on the same analyzer. In regression, the equation for the straight line is recast as y = bx + a. This change in terminology leads to confusion. Here a is the y-intercept or constant and b is the coefficient or slope of the line. A few more words of caution about regression - as in all of statistics there are certain assumptions: the x value is a true measure, both X and Y distributions are normal, and homoscedasticity, i.e., the variance of y is the same for each value of x. In this study the linearity classification made by different scientists were always in agreement. Typical examples of curves that were considered linear are presented in Fig. 1-5. Because these automated procedures values were usually within five percent of each other the curvature could be easily detected. The plot of the WBC, RBC, hemoglobin, hematocrit and platelet concentrations from approximately 74.4 to