A Study of the Effects of Use upon RGP Contact Lens Surface Ultrastructure

RGP Contact Lens 표면 미세구조에 대한 사용 효과의 연구

  • 김덕훈 (마산대학 안경광학과) ;
  • 성아영 (대불대학교 안경광학과) ;
  • Received : 2005.11.28
  • Published : 2006.01.31

Abstract

The aim of this study was to analyze the surface ultrastructure of RGP contact lenses post-use The RGP contact lenses, to investigate, were selected after one month, six months and one year of wear by the subjects. The anterior and posterior surfaces of these RGP contact lenses were inspected by a scanning electron microscope. The results were the following: 1. After one month of RGP contact lens use the anterior and posterior surfaces appeared clean and clear as originally. 2. After six months the anterior surface had several scars, dips, cracks and scratches. Upon the posterior surface appeared several foreign bodies and microorganisms. 3. After one year there was large scale damage and many foreign bodies were observed. Therefore, extended use of RGP contact lenses has shown increasing physical damage and extensive foreign body accumulation upon the ultrastructure of the lens surface.

본 연구의 목적은 사용한 RGP 콘택트렌즈의 표면 미세구조의 분석을 한 것이다. 조사를 위해 RGP콘택트렌즈는 피검자에 의해 1개월, 6개월 그리고 1년 사용한 것을 선택하였다. 이러한 RGP 콘택트렌즈의 전후 표면은 주사전자현미경으로 조사를 하였다. 그 결과는 다음과 같다. 1. RGP 콘택트렌즈 사용 1개월 후 전면과 후면의 표면은 선명하고 깨끗하였다. 2. 6개월 사용 렌즈의 전방 표면은 심한 반흔, 상처 자국, 파손 그리고 긁힘을 가지고 있었다. 그러나 후방 표면은 심한 이물질과 미생물 등이 나타났다. 3. 1년 사용렌즈는 좀더 심한 큰 반흔 손상과 많은 이물질들이 관찰되었다. 따라서 RGP 콘택트렌즈의 연장착용은 렌즈표면의 미세구조의 증가된 물리적인 손상과 심한 이물질의 누적을 나타내었다.