요부 내측지 고주파 신경절리술 시행 시 유도침 경로를 따라 발생한 화상 -증례보고-

Burn Wound along the Guide Needle Trajectory as a Complication of Radiofrequency Neurotomy of the Lumbar Medial Branch -A case report-

  • 윤홍준 (한양대학교 의과대학 마취통증의학교실) ;
  • 심재철 (한양대학교 의과대학 마취통증의학교실)
  • Youn, Hong Jun (Department of Anesthesiology and Pain Medicine, College of Medicine, Hanyang University) ;
  • Shim, Jae Chol (Department of Anesthesiology and Pain Medicine, College of Medicine, Hanyang University)
  • 투고 : 2006.08.21
  • 심사 : 2006.09.11
  • 발행 : 2006.12.30

초록

Radiofrequency neurotomy of the lumbar medial branch, via a caudal approach, is a representative interventional procedure for lumbar zygapophysial joint pain, which can be performed more accurately and easily using a guide needle technique. We experienced a case of burn wound formation along the guide needle trajectory, where heat conduction through the guide needle was suggested to have resulted in the burn wound.

키워드

참고문헌

  1. Kline MT, Yin W: Inrervenrional pain managemenr, 2nd ed. Philadelphia, W.B. Saunders. 2001, pp 243-93
  2. Sluijter ME: Radiofrequency. part 1: the lumbosacral region. Meggen (LU), Flivo Press SA. 2001, pp 49-72
  3. Ferrante FM, King LF, Roche EA, Kim PS, Aranda M, Delaney LR, et al: Radiofrequency sacroiliac joinr denervation for sacroiliac syndrome. Reg Anesth Pain Med 2001; 26: 137-42 https://doi.org/10.1097/00115550-200103000-00008
  4. Nakada SY, Jerde TJ, Warner TF, Wright AS, Haemmerich D, Mahvi DM, et al: Bipolar radiofrequency ablation of the kidney: comparison with monopolar radiofrequency ablation. J Endourol 2003; 17: 927-33 https://doi.org/10.1089/089277903772036316
  5. Anfinsen OG, Kongsgaard E, Foerster A, Aass H, Amlie JP: Radiofrequency currenr ablation of porcine right atrium: increased lesion size with bipolar two catheter technique compared to unipolar application in vitro and in vivo. Pacing Clin Electtophysiol 1998; 21:69-78 https://doi.org/10.1111/j.1540-8159.1998.tb01063.x
  6. Wang YG, Lu ZY, Zhao HY, Song YE, Li RL: A comparative study of radiofrequency ablation in unipolar and bipolar fashion. J Tongji Med Univ 1995; 15: 73-6 https://doi.org/10.1007/BF02887905
  7. Lee JH, Kim DW, Shim WS: Lumbar sympathetic radiofrequency thermocoagulation using bipolar probe in the hyperhidrosis patient. Korean J Pain 2005; 18: 92-5 https://doi.org/10.3344/kjp.2005.18.1.92
  8. Sim WS, Lee AR: Radiofrequency lumbar medial branch denervation using bipolar probe in patienr with facet joinr syndrome. Korean J Pain 2004; 17: 153-8 https://doi.org/10.3344/jkps.2004.17.2.153
  9. Steinke K, Gananadha S, King J, Zhao J, Morris DL: Dispersive pad site burns with modern radiofrequency ablation equipment. Surg Laparosc Endosc Percutan Tech 2003; 13: 366-71 https://doi.org/10.1097/00129689-200312000-00003
  10. Yagihara M, Takahashi H, Miyabe M, Toyooka H: [Skin burn after radiofrequency liver ablation of metastatic tumor]. Masui 2002; 51:1117-9
  11. Richtmyer JM: Elecrrosurgical burns in pediatric patients undergoing liver resection with saline-enhanced tadiofrequency technology. AORN J 2006; 83: 658-64; quiz 665-7
  12. Goette A, Reek S, Klein HU, Geller JC: Case report: severe skin burn at the site of the indifferent electrode after radiofrequency catheter ablation of typical atrial flutrer. J Interv Card Electrophysiol 2001; 5: 337-40 https://doi.org/10.1023/A:1011481104053