The purpose of this work is to predict the systolic blood pressure (BP) during exercise from pulse transit time (PTT) for warning of possible danger. PTT was calculated as the time between R-peak of ECG and the peak of differential photoplethysmograph (PPG). For the PTT-BP model, we used regress equations from previous studies and 3 kinds of new models combining linear and nonlinear regress equation. The model parameters were estimated with the data measured under low to middle intensity exercise, and then was tested with the data measured under high intensity exercise. Predicted BP values after high intensity exercise were compared with those measured by cuff-type sphygmomanometer. The results showed that the error between measured and predicted values were acceptable for the monitoring BP. We tested PTT-BP models 1 month after the identification without further calibration. Models could predict the BP and the errors between measured and predicted BP were about 5mmHg. The suggested system is expected to be helpful in recognizing any danger during exercise.

Keywords

References

  1. American College of Sports Medicine, ACSM's Guidelines for Exercise Testing and Prescription $6^{th}$ edition, Lippincott Williams & Wilkins, 2003, pp. 27-29
  2. Sheriwood, L., Human physiology: from cells to systems $5^{th}$ edition, Thomson Learning, 2004, pp. 302
  3. Pollak, M. H., and Obrist, P. A., 'Aortic-radial pulse transit time and ECG Q-wave to radial pulse wave as indices of beat-to-beat blood pressure change,' Psychophysiology, vol. 20,pp. 21-28,1983 https://doi.org/10.1111/j.1469-8986.1983.tb00895.x
  4. Kalju Meigas, Rain Kattai and Jaanus Lass, 'Continuous blood pressure monitoring using pulse wave delay,' in Proe. the 23rd Annual EMBS international Conference, Istanbul, Turkey, 2001, pp.3171-3174
  5. Smith, R. P., Argod, J., Pepin J. L. and Levy, P. A., 'Pulse transit time: and apprasial of potential clinical applications,' Thorax, vol. 54,pp.452-458, 1999 https://doi.org/10.1136/thx.54.5.452
  6. Paulus, D. A., 'Noninvasive blood pressure measurement,' Medical instrumentation, vol. 15, pp. 91-94, 1981
  7. C. M. Lee and Y. T. Zhang, 'Cuffless and noninvasive estimation of blood pressure based on a wavelet transform approach,' in Proc. IEEE EMBS Asian-Pacific Conference on Biomedical Engineering, Keihanna, Japan, 2003, pp. 148-149
  8. Parry Fung, Guy Dumon, Craig Ries, Chris Mott, Mark Ansermino, 'Continuous Noninvasive Blood Pressure Measurement by Pulse Transit Time,' in Proc. of the 26th Annual International Conference of the IEEE EMBS, C.A, USA, 2004, pp. 738-741
  9. Park, E. K., Cho, B. H., Park, S. H., Lee, J. y., Hwang, J. S., Park, H. K., Lee, J. S., Kim. I. Y. and Kim, S. I., 'A study on the estimation of continuous blood pressure using PIT and biometric parameters,' J Biomed. Eng. Res., vol. 27, pp. 1-5, 2006 https://doi.org/10.2220/biomedres.27.1
  10. Chen, W., Kobayashi, T., Ichikawa, S., Takeuchi, Y., and Togawa, T., 'Continuous estimation of systolic blood pressure using the pulse arrival time and intermittent calibration,' Med Biol Eng Comput, vol. 38, pp. 569-574,2000 https://doi.org/10.1007/BF02345755
  11. Jang, J. S. R., Sun, C. T., and Mizutani, E., Neuro-fuzzy and soft computing: a computational approach to learning and machine intelligence, Prentice-Hall, pp-104, 1997
  12. Association for the Advancement of Medical Instrumentation, Manual, electronic, or automated sphygmomanometers, American National standard, ANSI/AAMI SP10:2002 & ANSI/AAMI SP10:2002/Al:2003
  13. Lee, S. M., Park, E. K., Kim, I. Y., Kim, S. I., 'An estimating method for systolic blood pressure by using pulse transit time and physical characteristic parameters,' The Institute of electronics engineers of Korea, vol. 42, no. 3, pp. 145-150,2005