Development of Quality Assurance Program for the On-board Imager Isocenter Accuracy with Gantry Rotation
갠트리 회전에 의한 온-보드 영상장치 회전중심점의 정도관리 프로그램 개발
- Cheong, Kwang-Ho (Department of Radiation Oncology, Hallym University College of Medicine) ;
- Cho, Byung-Chul (Department of Radiation Oncology, Hallym University College of Medicine) ;
- Kang, Sei-Kwon (Department of Radiation Oncology, Hallym University College of Medicine) ;
- Kim, Kyoung-Joo (Department of Radiation Oncology, Hallym University College of Medicine) ;
- Bae, Hoon-Sik (Department of Radiation Oncology, Hallym University College of Medicine) ;
- Suh, Tae-Suk (Department on Biomedical Engineering College of Medicine, The Catholic University of Korea)
- 정광호 (한림대학교 의과대학 방사선종양학교실) ;
- 조병철 (한림대학교 의과대학 방사선종양학교실) ;
- 강세권 (한림대학교 의과대학 방사선종양학교실) ;
- 김경주 (한림대학교 의과대학 방사선종양학교실) ;
- 배훈식 (한림대학교 의과대학 방사선종양학교실) ;
- 서태석 (가톨릭대학교 의과대학 의공학교실)
- Published : 2006.12.31
Abstract
Positional accuracy of the on-board imager (OBI) isocenter with gantry rotation was presented in this paper. Three different type of automatic evaluation methods of discrepancies between therapeutic and OBI isocenter using digital image processing techniques as well as a procedure stated in the customer acceptance procedure (CAP) were applied to check OBI isocenter migration trends. Two kinds of kV x-ray image set obtained at OBI source angle of
본 연구에서는 방사선치료용 선형가속기의 갠트리 회전에 따른 온-보드 영상장치(on-board imager, OBI)의 회전중심점의 위치 정확도 확인을 위해 제조사에서 제공된 고객인수시험절차서(customer accetpance procedure, CAP)상에서 명시된 방법을 비롯하여 OBI 선원 위치