Analysis of the Respiratory Motion Effects on Dose Distribution Using TLD Phantom
열형광선량계용 팬톰을 이용한 호흡 움직임에 따른 선량분포의 평가
- Hong, Ju-Young (Department of Biomedical Engineering, The Catholic University of Korea College of Medicine) ;
- Kim, Yon-Lae (Department of Radiation Oncology, Konkuk University Hospital) ;
- Rah, Jeong-Eun (Department of Biomedical Engineering, The Catholic University of Korea College of Medicine) ;
- Chung, Jin-Beom (Department of Biomedical Engineering, The Catholic University of Korea College of Medicine) ;
- Suh, Tae-Suk (Department of Biomedical Engineering, The Catholic University of Korea College of Medicine)
- 홍주영 (가톨릭대학교 의과대학 의공학교실) ;
- 김연래 (건국대학교병원 방사선종양학과) ;
- 라정은 (가톨릭대학교 의과대학 의공학교실) ;
- 정진범 (가톨릭대학교 의과대학 의공학교실) ;
- 서태석 (가톨릭대학교 의과대학 의공학교실)
- Published : 2006.12.31
Abstract
The purpose of this study was to measure the dose distribution from the moving phantom for the respiratory motion. The phantom for TLD measurement was designed and built for this study based on the multiple plates for placing TLD and film. The TLDs may be inserted at 3 mm intervals in each TLD plate. For the measurements, TLD plate was inserted into the phantom at 1.5 cm (
본 연구에서는 열형광선량계(TLD)용 팬톰을 설계 및 제작하고 구동장치에 위치시켜 팬톰이 호흡과 유사하게 움직이는 경우에 선량분포의 변화를 측정하였다. 제작된 팬톰은 TLD를 최소 3 mm 간격으로 위치시킬 수 있는 TLD용 판과 필름을 장착시킬 수 있는 판으로 구성된다. 선량분포 측정을 위하여 팬톰 속 깊이 1.5 cm (