Application of Pulse Pile-Up Correction Spectrum to the Library Least-Squares Method

펄스 중첩 보정 스펙트럼의 라이브러리 최소자승법에의 이용

  • Lee, Sang-Hoon (Department of Physics Kyungpook National University)
  • Published : 2006.12.30

Abstract

The Monte Carlo simulation code CEARPPU has been developed and updated to provide pulse pile-up correction spectra for high counting rate cases. For neutron activation analysis, CEARPPU correction spectra were used in library least-squares method to give better isotopic activity results than the convention library least-squares fitting with uncorrected spectra.

개선된 몬테칼로 코드 CEARPPU는 고계수율 상황에서의 펄스 중첩 보정 스펙트럼을 제공한다. 중성자 방사화 분석을 위하여 CEARPPU를 이용하여 보정된 스펙트럼으로 라이브러리 최소자승 동위원소 방사능 분석을 수행하여 보정하지 않은 스펙트럼을 이용하는 방법보다 우수한 결과를 얻었다.

Keywords

References

  1. Ao Q, Lee SH, and Gardner RP. Error analysis of the in vivo XRF measurement of lead in bone by Monte Carlo simulation. Advances in X-ray Analysis 1999
  2. Todd AC, McNeill FE, and Fowler BA. In vivo X-ray fluorescence of lead in bone. Environ. Res. 59:326-335 https://doi.org/10.1016/S0013-9351(05)80039-8
  3. Preiss IL and Washington II W. Skin thickness effects on in vivo LXRF. Adv. X-ray Anal. 1995:38:607-613
  4. Gardner RP, Lee SH, Guo W, et al., CEARPPU: A Monte Carlo code for predicting pulse pile-up distortion for high counting-rate radiation spectra, PSR-528. Radiation Safety Information Computational Center, Oak Ridge National Laboratory, 2004
  5. Wielopolski L and Gardner RP. Prediction of the pulse-height spectral distortion caused by the peak pile-up effect Nucl. Instr. and Meth. 1976;133:303-309 https://doi.org/10.1016/0029-554X(76)90623-6
  6. Gardner RP and Wielopolski L. A generalized method for correcting pulse- height spectra for the peak pile-up effect due to double sum pulses: Part I. Predicting spectral distortion for arbitrary pulse shapes. Nucl. Instr. and Meth. 1977;140:289-296 https://doi.org/10.1016/0029-554X(77)90297-X
  7. Wielopolski L and Gardner RP. A generalized method for correcting pulse- height spectra for the peak pile-up effect due to double sum pulses: Part II. The inverse calculation for obtaining true from observed spectra. Nucl. Instr. and Meth. 1977;140:297-303 https://doi.org/10.1016/0029-554X(77)90298-1
  8. Bristow Q and Harrison RG. Theoretical and experimental investigations of the distortion in radiation spectra caused by pulse pileup. Nucl. Geophys. 1991;5:141-186
  9. ORTEC, Model 572 Spectroscopy Amplifier and Pile-Up Rejector Operating and Service Manual. 1982
  10. Knoll GF. Radiation Detection and Measurement. New York; Wiley, 1979