Comparison and Analysis of Photon Beam Data for Hospitals in Korea and Data for Quality Assurance of Treatment Planning System
국내 의료기관들의 광자 빔 데이터의 비교 분석 및 치료계획 시스템 정도관리자료
- Lee, Re-Na (Department of Radiation Oncology, Ewha Womans University College of Medicine) ;
- Cho, Byung-Chul (Department of Radiation Oncology, Hallym University College of Medicine) ;
- Kang, Sei-Kwon (Department of Radiation Oncology, Hallym University College of Medicine)
- Published : 2006.09.30
Abstract
Purpose: Photon beam data of linear accelerators in Korea are collected, analyzed, and a simple method for checking and verifying the dose calculations in a TPS are suggested. Materials and Methods: Photon beam data such as output calibration condition, output factor, wedge factor, percent depth dose, beam profile, and beam quality were collected from 26 institutions in Korea. In order to verify the accuracy of dose calculation, ten sample planning tests were peformed. These Include square, elongated, and blocked fields, wedge fields, off-axis dose calculation, SSD variation. The planned data were compared to that of manual calculations. Results: The average and standard deviation of photon beam quality for 6, 10, and 15 MV were
목적: 방사선 종양학과에서 사용되고 있는 선형가속기의 광자선 빔 데이터를 수집하여 비교 분석하였으며 치료계획용 시스템에 대한 간단한 정도관리 방법을 제시하였다. 대상 및 방법: 국내 26개 방사선 치료기관을 대상으로 출력교정 조건, 출력인자, 쐐기인자, 깊이 선량분포, 측방선량분포 및 선질에 대한 데이터를 수집하였다. 치료계획용 시스템의 선량계산의 정확성을 확인하기 위하여 10가지 광자선 치료 조건(정방형/직사각형/부정형 조사면, 쐐기필터 조사면, 축이탈 선량계산, SSD 변화)에 대한 선량계산을 치료계획용 시스템을 이용하여 시행하였으며 치료계획용 시스템을 이용하여 계산된 모니터 값과 수 계산에 의한 결과를 비교 분석하였다. 결과: 광자선 선질은 6 MV, 10 MV 및 15 MV에 대해 각각
Keywords