Dosimetric Characterization of an Ion Chamber Matrix for Intensity Modulated Radiation Therapy Quality Assurance
세기변조방사선치료 선량분포 확인을 위한 2차원적 이온전리함 배열의 특성분석
- Lee, Jeong-Woo (Department of Radiation Oncology, Konkuk University Hospital) ;
- Hong, Se-Mie (Department of Radiation Oncology, Konkuk University Hospital) ;
- Kim, Yon-Lae (Department of Radiation Oncology, Konkuk University Hospital) ;
- Choi, Kyoung-Sik (Department of Radiation Oncology, Ajou University Hospital) ;
- Jung, Jin-Beom (Seoul National University Bundang Hospital) ;
- Lee, Doo-Hyun (National Cancer Center) ;
- Suh, Tae-Suk (Biomedical Engineering, School of Medicine, The Catholic University of Korea)
- 이정우 (건국대학교병원 방사선종양학과) ;
- 홍세미 (건국대학교병원 방사선종양학과) ;
- 김연래 (건국대학교병원 방사선종양학과) ;
- 최경식 (아주대학교병원 방사선종양학과) ;
- 정진범 (분당서울대학교병원 방사선종양학과) ;
- 이두현 (국립암센터 방사선종양학과) ;
- 서태석 (가톨릭대학교 의과대학 의공학교실)
- Published : 2006.09.30
Abstract
A commercial ion chamber matrix was examined the characteristics and its performance for radiotherapy qualify assurance. The device was the I'mRT 2D-MatriXX (Scanditronix-Wellhofer, Schwarzenbruck, Germany). The 2D-MatriXX device consists of a 1020 vented ion chamber array, arranged in
본 연구의 목적은 2차원적 이온전리함 배열의 특성을 분석하고 세기변조방사선치료 선량분포 확인의 유용성을 평가하기 위함이다. 본 실험에 사용된 2차원적 이온전리함 배열(MatriXX, Scanditronix-Wellhofer, IBA, Germany)은