Biomechanical Finite Element Analysis of Bone Cemented Hip Crack Initiation According to Stem Design

  • Kim, Byeong-Soo (Department of Mechanical and Atomobile Engineering, Inje University) ;
  • Moon, Byung-Young (Industrial Liaison Innovation Cluster, Pusan National University) ;
  • Park, Jung-Hong (Department of Mechanical Design Engineering, Pusan National University)
  • 발행 : 2006.12.01

초록

The purpose of this investigation was to determine the specific fracture mechanics response of cracks that initiate at the stem-cement interface and propagate into the cement mantle. Two-dimensional finite element models of idealized stem-cement-bone cross-sections from the proximal femur were developed for this study. Two general stem types were considered; Rectangular shape and Charnley type stem designs. The FE results showed that the highest principal stress in the cement mantle for each case occurred in the upper left and lower right regions adjacent to the stem-cement interface. There was also a general decrease in maximum tensile stress with increasing cement mantle thickness for both Rectangular and Charnley-type stem designs. The cement thickness is found to be one of the important fatigue failure parameters which affect the longevity of cemented femoral components, in which the thinner cement was significantly associated with early mechanical failure for shot-time period.

키워드

참고문헌

  1. ?Barsom, J. M. and Rolfe, S. T., 1987, Fracture and Fatigue Control in Structures, 2nd Ed., Prentice-Hall
  2. Bittencourt, T. N., 1993, 'Computer Simulation of Linear and Nonlinear Crack Propagation in Cementitious Materials,' Ph D Thesis, Cornell University
  3. Crowninshield, R. D., Brand, R. A., Johnston, R. C. and Milroy, J. C., 1980, 'The Effect of Femoral Cross-sectional Geometry on Cement Stresses in Total Hip Reconstruction,' Clinical Orthopedics, Vol. 146, pp. 71-77
  4. Ebramzadeh, E., Sarmiento A., McKellop H. A., Llinas A. and Gogan W., 1994, 'The Cement Mantle in Total Hip Arthroplasty. Analysis of Long-Term Radiographic Results,' Journal of Bone and Joint Surgery, Vol. 76(A), pp. 77-87
  5. Estok, D. M., Orr, T. E. and Harris, W. H., 1997, 'Factors Affecting Cement Strains Near the Tip of a Cemented Femoral Component,' The Journal of Arthroplasty, Vol. 12, No. 1, pp.40-48 https://doi.org/10.1016/S0883-5403(97)90045-0
  6. James J. M. and Ares J. R., 1993, 'On the Dependence of the Dynamic Crack Tip Temperature Fields in Metals Upon Crack Tip Velocity and Material Parameters,' Mechanics of Materials, Vol. 16, No. 4, pp. 337-350 https://doi.org/10.1016/0167-6636(93)90009-G
  7. Jasty, M., Maloney, W. J., Bragdon, C. R., Haire, T. and Harris, W. H., 1990, 'Histomorphological Studies of the Long-Term Skeletal Re?sponses to well Fixed Cemented Femoral Components,' Journal of Bone Joint Surgery, Vol. 72 (A), pp. 1220-1229 https://doi.org/10.2106/00004623-199072080-00015
  8. Kawate, K., Maloney, W. J., Bragdon, C. R., Biggs, S. A., Jasty, M. and Harris, W. H., 1998, 'Importance of a Thin Cement Mantle. Autopsy Studies of Eight Hips,' Clinical Orthopedic Related Research, Vol. 355, pp. 70-76 https://doi.org/10.1097/00003086-199810000-00008
  9. Kim, B. S. and Mann, K. A., 1999, 'A Thin Cement Mantle Decreases Fatigue Life of Cemented Femoral Hip Component,' Transactions of Orthopaedic Research Society, Vol. 24, pp. 880
  10. Leis, B. N., Hopper, A. T., Broek, J. A. and Kanninen, M. F., 1986, 'Critical Review of the Fatigue Growth of Short Cracks,' Engineering Fracture Mechanics, Vol. 23, No.5, pp. 883-898 https://doi.org/10.1016/0013-7944(86)90099-8
  11. Maloney, W., Jasty, M., Rosenburg, A. and Harris, W. H., 1990, 'Bone Lysis in Well Fixed Femoral Commponents,' Journal of Bone and Joint Surgery, Vol. 72(B), pp. 966-970
  12. Mann, K. A., Bartel, D. L. and Ayers, D. C., 1997, 'Influence of Stem Geometry on Mechanics of Cemented Femoral Hip Components With a Proximal Bond,' Journal of Orthopaedic Research, Vol. 15, No. 5, pp.700-706 https://doi.org/10.1002/jor.1100150511
  13. Mann, K. A., Bartel, D. A., Wright, T. M. and ?Ingraffea, A. R., 1991, 'Mechanical Characteristics of the Stem-cement Interface,' Journal of Orthopaedic Research, Vol. 9, pp. 798-808 https://doi.org/10.1002/jor.1100090605
  14. McCormack, B. A. O., Prendergast, P. J. and Gallagher, D. G., 1996, 'An Experimental Study of Damage Accumulation in Cemented Hip Prostheses,' Clinical Biomechanics, Vol. 11, No.4, pp.214-219 https://doi.org/10.1016/0268-0033(95)00076-3
  15. Mulroy, R. D. and Harris, W. H., 1990, 'The Effect of Improved Cementing Techniques on Component Loosening in Total Hip Replacements: an 11-year Radiographic Review,' Journal of Bone and Joint Surgery, Vol. 72 (B) , pp. 757-760
  16. Rimnac, C. M., Wright, T. M. and McGill, D. L., 1986, 'The Effect of Centrifugation on the Fracture Properties of Acrylic Bone Cements,' Journal of Bone and Joint Surgery, Vol. 68 (A) , pp. 281-287
  17. Star, M. J., Colwell, C. W., Kelman, G. J., Ballock, R. T. and Walker, R. H., 1994, 'Suboptimal (thin) Distal Cement Mantle Thickness as a Contributory Factor in Total Hip Arthroplasty Femoral Component Failure: A Retrospective Radiographic Analysis Favoring Distal Stem Centralization,' The Journal of Arthroplasty, Vol. 9, No. 2, pp. 143-149 https://doi.org/10.1016/0883-5403(94)90063-9