A noble RBC aggregometer with vibration-induced disaggregation mechanism

  • Shin S. (School of Mechanical Engineering, Kyungpook National University) ;
  • Jang J.H. (School of Mechanical Engineering, Kyungpook National University) ;
  • Park M.S. (School of Mechanical Engineering, Kyungpook National University) ;
  • Ku Y.H. (School of Mechanical Engineering, Kyungpook National University) ;
  • Suh J.S. (Dept. of Laboratory Medicine, School of Medicine, Kyungpook National University Hospital)
  • Published : 2005.03.01

Abstract

The aggregation of red blood cells (RBCs) is a major determinant of blood flow resistance passing through various veins. Available techniques for measuring RBC aggregation often require pretreating and washing after each measurement, which is not optimal for day-to-day clinical use. A laser reflection technique has been combined with a vibration-aided disaggregation mechanism, which shows significant advances in aggregometer design, operation and data analysis. The essential features of this design are in its simplicity and a disposable element that is in contact with the blood sample. Using extremely small quantities of blood, the RBCs subjected to vibrations can be quickly and completely disaggregated. This is followed by measuring the backscattered light intensity. The measurements with the present sensor were compared with those of a commercial aggregometer and a strong correlation was found between them. The newly-developed optical aggregometer can measure the RBC aggregability difference between young and old cell suspension with ease and accuracy.

Keywords

References

  1. Baskurt, O. K., A. Temiz and H. J. Meiselman, 1997, Red blood cell aggregation in experimental sepsis, J. Lab. Clin. Med. 130, 183 https://doi.org/10.1016/S0022-2143(97)90094-9
  2. Bauersachs, R. M., S. Shaw, A. Zeidler and H. J. Meiselman, 1987, Hemorheological finding in uncontrolled type II diabetes mellitus, Clin. Hemorheol. 7, 432
  3. Boynard, M., J.C. Lelievre and R. Guillet, 1987, Aggragtionof red blood cells studied by ultrasound back-scattering phenomenon, Biorheology 24, 451 https://doi.org/10.3233/BIR-1987-24503
  4. Chien, S. and K. M. Jan, 1973, Ultrastructural basis of the mechanism of the rouleaux formation, Microvas. Res. 5, 155-166 https://doi.org/10.1016/0026-2862(73)90068-X
  5. Dormandy, L., E. Ernst, A. Matrai and P.T. Flute, 1982, Hemorheologic changes following acute myocardial infarction, Am. Heart J. 104, 1364 https://doi.org/10.1016/0002-8703(82)90169-7
  6. Hardeman, M.R., P.T. Goedhart, J.G.G. Dobbe and K.P. Lettinga, 1994, Laser-assisted optical rotational cell analyser (L.O.R.C.A); I. A new instrument for measurement of various structural hemorheological parameters, Clinical Hemorheology 14, 605-618
  7. Hein, H. J., R. M. Bauersachs, E. I. Feinstein and H. J. Meiselman, 1987, Hemorheological abnormalities in chronic renal failure patients, Clin. Hemorheol. 7, 430
  8. Meiselman, H. J., O. K. Baskurt, S. O. Sowemimo-Coker and R. B. Wenby, 1999, Cell electrophoresis studies relevant to red blood cell aggregation, Biorheology. 36, 427
  9. Neu, B. and H. J. Meiselman, 2002, Depletion-mediated red blood cell aggregation in polymer solutions, Biophys. J. 83, 2482 https://doi.org/10.1016/S0006-3495(02)75259-4
  10. Pribush, A., H.J. Meiselman, D. Meyerstein and N. Meyerstein, 1999, Dielectric approach to the investigation of erythrocyte aggregation : I. Experimental basis of the method, Biorheology 36, 411
  11. Rampling, M.W., 1988, Red cell aggregation and yield stress, in Clinical Blood Rheology Vol. 1, edited by G.D. Lowe, CRC Press, Boca Raton, Florida, USA
  12. Rampling, M. W., H. J. Meiselman, B. Neu and O. K. Baskurt, 2004, Influence of cell-specific factors on red blood cell aggregation, Biorheology 41, 91
  13. Shvartsman, L.D. and I. Fine, 2001, Light scattering changes caused by RBC aggregation : physical basis for new approach to non-invasive blood count, SPIE proc. 4263, 122
  14. Shin, S., Y.H. Ku, J.S. Suh, S.Y. Moon and J.Y. Jang, 2003, Characteristics of blood flow resistance under transverse vibration : red blood cell suspension in Dextran-40, Annals of Biomedical Engineering 31, 1077 https://doi.org/10.1114/1.1603751
  15. Shin, S., M.S. Park, J.H. Jang, Y.H. Ku and J.S. Suh, 2004, Measurement of red blood cell aggregation by analysis of light transmission in a pressure-driven slit flow system, Korea-Austrailia Rheology Journal 16, 129
  16. Shin, S., J.H. Jang, M.S. Park, Y.H. Ku and J.S. Suh, 2005, Light-transmission aggregometer using a vibration-induced disaggregation mechanism, Review of Scientific Instrum. 76, 016107 1 https://doi.org/10.1063/1.1832456
  17. Stoltz, J. F., M. Singh and P. Riha, 1999, Hemorheology in Practice. IOS Press, Amsterdam
  18. Zhao, H., X. Wang and J.F. Stolz, 1999, Comparison of three optical methods to study erythrocyte aggregation, Clinical Hemorheol. And Microcirc. 21, 297