A study on Phantom image of Mammography in some area

일부지역 유방촬영술의 팬톰영상에 관한 연구

  • Kang, Se-Sik (Dept. of Radiological Science, College of Health science, Catholic University of Busan)
  • 강세식 (부산가톨릭대학교 보건과학대학 방사선학과)
  • Published : 2005.12.30

Abstract

Purpose : The goal of this study was to survey the quality control of mammographic phantom images in some area. Methods : Between May and June, 2005, I have got a total of 47 mammographic phantom images collected from 49 hospitals and were analized about phantom score of fibrous, speck, tumor with enlargement scope. By critic level of experimental mammographic phantom case images, classified through the pass case that fibrous score 4, speck score 3, tumor score 3 over. Result : It passed 33case (70.3%), failed 14 case (29.7%) about 47 case. Passing case images showed tube voltage $23{\sim}30\;kVp$, tube current $20{\sim}60\;mAs$, photographic density $0.48{\sim}1.77$ Conclusion : It passed case about 70.3% of mammographic phantom image in some area. Optimum range of tube voltage, tube current, photographic density have to use to get fine quality of mammographic images.

목 적 : 일부지역에서 실시되고 있는 유방촬영술에 대한 유방팬톰영상을 분석하여 화질관리의 실태를 알아보고져 한다. 대상과 방법 : 2005년 5월에서 6월까지 2개월간 일부지역의 47개 병원에 설치된 유방촬영기기를 대상으로 실험하였으며 유방팬톰영상의 분석은 확대경을 이용하여 영상에서 보이는 섬유소, 작은 석회화그룹, 종괴의 개수를 세어 점수로 기록 평가하였다. 유방팬톰영상평가시험의 평가기준에 의해서 섬유소 4개 이상, 작은 석회화그룹 3개 이상, 종괴 3개 이상이 보인 경우를 합격으로 분류하였다. 결 과 : 유방팬톰영상 47예 중 합격이 33예(70.3%), 불합격이 14예(29.7%)였으며 합격한 경우에 사용한 관전압 범위는 $23{\sim}30\;kVp$, 관전류량 범위는 $20{\sim}60\;mAs$, 사진농도 범위는 $0.48{\sim}1.77$로 나타났으며 격자를 사용한 경우는 31예(91%)로 나타났다. 결 론 : 일부지역에서 실시되고 있는 유방촬영술의 영상에서 약 70.3%가 유방팬톰영상평가 시험에 합격한 것으로 나타났다. 적절한 범위의 관전압, 관전류, 격자 사용으로 적절한 사진농도를 얻어 시험기준에 합격된 우수한 화질의 유방영상을 얻어야 한다.