The Effect of the CT Number for Each CT on Dose Calculation
CT 기종에 따른 CT 수의 변화가 선량계산에 미치는 영향
- Cho Kwang Hwan (Department of Radiation Oncology, College of Medicine, Soonchunhyang University, Department of Medical Physics, Kyonggi University) ;
- Lee Suk (Department of Radiation Oncology, College of Medicine, Korea University) ;
- Cho Sam Ju (Department of Medical Physics, Kyonggi University) ;
- Lim Sangwook (Department of Medical Physics, Kyonggi University) ;
- Huh Hyun Do (Department of Medical Physics, Kyonggi University) ;
- Min Chul Kee (Department of Medical Physics, Kyonggi University) ;
- Cho Byung-Chul (Department of Radiation Oncology, Hallym University College of Medicine) ;
- Kim Yong Ho (Department of Radiation Oncology, College of Medicine, Soonchunhyang University) ;
- Choi Doo Ho (Department of Radiation Oncology, College of Medicine, Soonchunhyang University) ;
- Kim Eun Seog (Department of Radiation Oncology, College of Medicine, Soonchunhyang University) ;
- Kwon Soo Il (Department of Medical Physics, Kyonggi University)
- 조광환 (순천향대학교 의과대학 방사선종양학교실, 경기대학교 의학물리학과) ;
- 이석 (고려대학교 의과대학 방사선종양학교실) ;
- 조삼주 (경기대학교 의학물리학과) ;
- 임상욱 (경기대학교 의학물리학과) ;
- 허현도 (경기대학교 의학물리학과) ;
- 민철기 (경기대학교 의학물리학과) ;
- 조병철 (한림대학교 의과대학 방사선종양학교실) ;
- 김용호 (순천향대학교 의과대학 방사선종양학교실) ;
- 최두호 (순천향대학교 의과대학 방사선종양학교실) ;
- 김은석 (순천향대학교 의과대학 방사선종양학교실) ;
- 권수일 (경기대학교 의학물리학과)
- Published : 2005.12.01
Abstract
The CT number corresponds to electron density and its influence on dose calculation was studied. Five kinds of CT scanners were used to obtain Images of electron density calibration phantom (Gammex RMI 467), Then the differences between CT numbers for each scanners were
CT 기종에 따라서 전자밀도에 대응하는 CT 수의 변화와 치료계획 시 선량계산에 미치는 영향을 분석하였다. 5종의 CT를 이용하여 전자밀도 교정 팬톰의 영상을 얻어 기종에 따른 CT 수의 변화를 알아보았다. 조직등가물질의 밀도에서 CT 수는