Comparison of Magnetocardiogram Parameters Between a Ischemic Heart Disease Group and Control Group

정상군 및 허혈성 심질환 환자군에서의 심자도 파라미터 비교

  • 박종덕 (한국전기연구원 영상응용연구그룹) ;
  • 허영 (한국전기연구원 영상응용연구그룹) ;
  • 진승오 (한국전기연구원 영상응용연구그룹) ;
  • 전성채 (한국전기연구원 영상응용연구그룹)
  • Published : 2005.11.01

Abstract

The electrical current generated by heart creates not only electric potential but also a magnetic field. We have observed electrophysiological phenomena of the heart by measuring components of magnetocardiogram(MCG) using 61 channel superconducting quantum interference device(SQUD) system. We have analyzed the possibility and characteristics of MCG parameters for diagnosis of ischemic heart disease. A technique for automatic analysis of MCG signals in time domain was developed. The methods for detecting the position, the interval, the amplitude ratio, and the direction of single current dipole were examined in the MCG wave. The position and interval parameters were obtained by calculating the gradients of a envelope curve which could be formed by the difference between the maximum and minimum envelope of multi-channel MCG signals. We show some differences of the frequency contour map between the normal MCG and the abnormal (ischemic heart disease) MCG. The direction of single current dipole can be defined by rotating the magnetic field according to Biot-Savart's law at each point of MCG signals. In this study, we have examined the direction of single current dipole from searching for the centroids of positive and negative magnetic fields. The amplitude ratio parameters for measuring 57 deviation consisted of A$_{T}$/A$_{R}$ and other ratios. and We developed a new analysis method, which is based on the frequency contour map of electromagnetic field. Using theses parameters, we founded significant differences between normal subjects and ischemic patients in some parameters.

Keywords

References

  1. Kim K, Lee YH, Kwon H, Kim JM, Park YK, 'Detector Array Optimization for an MCG System Measuring Tangential Field : A favorable comparison with normal MCG measurement', Biomag2004, pp. 359-360, 2004
  2. Williamson SJ, Hoke M, Kotani 'Advances in Biomagnetism Plenum Press', New York. pp.555-558, 1989
  3. L.Link, P.Endt, U.Steinhoff, P.Goedde and L.Trahms, 'Analysis of QRS shape variability and short-term heart rate variability of CAD patients' Computers in Cardiology pp. 13-16, Sept. 1998 https://doi.org/10.1109/CIC.1998.731820
  4. Baule GM, McFee R 'Detection of the Magnetic Field of the Heart,' Am heart, 66:95-6, July 1963 https://doi.org/10.1016/0002-8703(63)90075-9
  5. C. Cohen-Tannoudji, J DuPont-Roc, S. Haroche, and F. Laloe, 'Detection of the Static Magnetic Field Produced by the Oriented Nuclei of Optically Pumped 3He Gas', Phys, Rev. Lett. 22, 758, 1969 https://doi.org/10.1103/PhysRevLett.22.758
  6. M. Sato, Y. Terada, T.Mitsui, T. Miyashita, A. Kandori and K. Tsukada, 'Detection of myocardial ischemia by magnetocardiogram using 64-channel SQUID system' Proceedings of 12th International Conference on Biomagnetism, pp. 523-526, Helsinki, Finland, 2000
  7. K. Okajima, T.Komuro, N.Harada, A.Adachi, M.Ueda, AKandori, G.Uehara and H.Kado, 'A 64-channel DC-SQUID magnetometer system' Nuclear Science Symposium and Medical Imaging Conference, vol.3, pp. 1494-1497 1993
  8. W.C. Chen, Q.S Yang, G.H. Chen, L.H Zhang, J. Feng, X.G. Huang, Y.P. Liu and Z.Q. Wang, 'Parametric analysis of MCG wave sets in Chinese' Proceedings of 13th International Conference on Biomagnetism, pp, 566-567, Jena, Germany, 2002
  9. J. Lotjonen, P.-J Reissrnan, I.E. Magnin, and T. Katila, 'Model extraction from magnetic resonance volume data using the deformable pyramid', Med. Image Anal. 3, 387-406, 1999 https://doi.org/10.1016/S1361-8415(99)80031-0
  10. Stephan Achenbach and Werner Moshage, 'Magnetocardiography: clinical investigations with a biomagnetic multichannel system', Physiol, Meas. 14, A61- A68, 1993 https://doi.org/10.1088/0967-3334/14/4A/011
  11. 전창익, 허영, 김기욱, 진승오, 장원석, 이현호, '다채널 심자도 신호의 파라메터 추출에 관한 연구', 제16회 신호처리합동학술대회, 2003
  12. Chang Ik Jeon, Young Huh, Byung Hee Han, Won Suk Chang, Seung Oh Jin, Doo Ho Youm, Ki Uk Kim, Hong Sik Byun, Sang Hoon Lee, Seok Y oon Choi, Sung Hee Choi, 'Automatic Extraction of the Diagnostic Parameters in Magnetocardiogram', the 2004 International Technical Conference on Circuit/Systems, Computers and Communication, pp. 6B2L -5, July 6-8, 2004
  13. M. Nomura, Y. Nakaya, M. Yukinaka, Y. Kondo, K. Saito, and S. Ito. 'QTc dispersion using magnetocardiogram: analysis in patients with myocardial infarction', J. Jpn. Biomagnetis. and Bioelectromagnetics Society 10, 96-97, 1997
  14. G. Eason, B. Noble, and I.N. Sneddon, 'Principles of Software QRS Detection.,' IEEE Eng. Med. Biol' Mag., vol.21, pp. 41-57, February 2002 https://doi.org/10.1109/51.993193
  15. Kim, K, Lee, Y. H., Kwon, H., Kim, J. M., Kim, I. S., Park, Y. K, 'Averaging Algorithm Based on Data Statistics in Magnetocardiography', Biomag2004, pp. 357 - 358, 2004
  16. 전창익, 허 영, 김기욱, 한병희, 진승오, 장원석, 이현호, '심자도 전류원 방향 추정과 허혈성 심질환 진단', 대한의용생체공학회 추계학술대회, pp. 126, 2003
  17. 허 영, 전호철, 박종덕, 진승오, 전창익, 변홍식 '심자도 주파수 평면 해석에 의한 혀혈성 심질환 진단에 관한 연구', 대한의용생체공학회 춘계학술대회, pp. 110, 2005