Reproducibility of panoramic radiography in patients

임상에서 촬영되는 파노라마 방사선사진의 재현성 조사

  • Nah Kyung-Soo (Department of Oral and Maxillofacial Radiology, College of Dentistry, Pusan National University)
  • 나경수 (부산대학교 치과대학 구강악안면방사선학교실)
  • Published : 2005.09.01

Abstract

Purpose : To evaluate the magnification error percentage in repeatedly taken panoramic radiographs of same patient and machine. Materials and Methods : 92 panoramic radiographs from 46 patients were traced and 30 horizontal and vertical measurements were made with digital sliding caliper. The results were compared with paired t-test. Results : There was no statistically significant difference between the two measurements. The overall difference as percentage error was $6.19\pm5.60\%$. The largest error as $14.61\pm12.44\%$ was found at condylar height 1, and smallest as $1.86\pm1.61\%$ at mandibular height. Overall vertical error excluding condylar height 1 was $3.76\pm3.97\%$, and the horizontal error $6.88\pm5.92\%$. Conclusion . Repeatedly taken panoramic radiographs of the same patient and machine was reliable since there was no significant percentage error difference but the percentage error ranged from $1.86\pm1.61\%\;to\;14.61\pm12.44\%$ indicating the error depends on the measuring site.

Keywords