The evaluation of usefulness of the newly manufactured immobilization device
치료보조기구의 제작 및 유용성 평가
- Seo Seok Jin (Dept. of Radiation Oncology, Seoul National University Hospital) ;
- Kim Chan Yoeng (Dept. of Radiation Oncology, Seoul National University Hospital) ;
- Lee Je Hee (Dept. of Radiation Oncology, Seoul National University Hospital) ;
- Park Heung Deuk (Dept. of Radiation Oncology, Seoul National University Hospital)
- Published : 2005.03.01
Abstract
Purpose : To evaluate the usefulness of the handmade patient immobilization device and to report the clinical results of it. Materials and methods : We made two fusion images and analyzed those images. One image is made with diagnostic MR image and CT image, the other with therapeutic planning MR image and CT image. With open head holder, we measured the skin dose and attenuation dose. Also, we made the planning CT couch plate with acrylic plate and styrofoam and compared artifact. Results : We could get more accurate fusion image when we use MR head holder(within 2mm error). The skin dose was reduced 2 times and the attenuation dose was reduced more than
목적 : 최근의 방사선치료는 정상조직의 장애는 최대한 줄이고 종양조직 선량을 높여 치료효과를 극대화 하기위하여 3차원적으로 발전되어왔으며, 이에 따른 환자 체위고정을 위한 고정기구와 보조기구의 개발이 필수적으로 요구되고 있다. 본원에서는 MR head holder, 개방형 head holder, Plan CT couch plate를 자체 제작하여 그 유용성을 평가하고 분석하였으며, 임상에 적용한 결과를 보고하고자 한다. 대상 및 방법 : MR head holder를 제작하여 진단 MR영상과 치료계획용 MR영상을 CT영상과 fusion하여 두 영상의 일치성과 정확성을 분석하였고, 개방형 head holder를 제작하여 피부선량과 선량감약 정도를 알아보았으며, 6 mm 아크릴판에 스티로폼을 부착하는 CT couch plate를 제작하여 hard plate와 artifact 정도를 비교하였다. 결과 : Head holder를 사용하지 않은 fusion영상보다 2 mm이내의 오차로 정확한 fusion을 할 수 있었고, 개방형 head holder를 사용함에 따라 피부선량을 최대 2배 이상 줄일 수 있었고,