Progress in Medical Physics (한국의학물리학회지:의학물리)
- Volume 16 Issue 3
- /
- Pages.125-129
- /
- 2005
- /
- 2508-4445(pISSN)
- /
- 2508-4453(eISSN)
The Comparison of Beam Data between Measured Beam Data and Calculated Beam Data Using Treatment Planning System
6 MV 광자선의 측정데이터와 치료계획장치에 의한 계산데이터의 비교
- Park Sung Kwang (Department of Radiation Oncology, College of Medicine, Inje University) ;
- Cho Byung Chul (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Cho Heung Lae (Department of Radiation Oncology, College of Medicine, Inje University) ;
- Ahn Ki Jung (Department of Radiation Oncology, College of Medicine, Inje University)
- 박성광 (인제대학교 의과대학 부산백병원 방사선종양학과) ;
- 조병철 (한림대학교 의과대학 방사선종양학교실) ;
- 조흥래 (인제대학교 의과대학 부산백병원 방사선종양학과) ;
- 안기정 (인제대학교 의과대학 부산백병원 방사선종양학과)
- Published : 2005.09.01
Abstract
The first step in the commissioning procedure of a treatment planning system is always verification of the basic beam data. In this work, we have measured POD curves and beam profiles between 1
입체조형방사선치료와 세기조절방사선치료를 위해 여러 셈법을 포함한 치료계획장치가 임상에 사용되고 있다. 본 연구에서는 조사면 1