Study for Multi Channel Radiation Detector Using of Microfilm and Carbon Electrode
탄소막 마이크로필름을 이용한 다채널 전리함 개발에 관한 연구
- Shin Kyo Chul (Department of Radiation Oncology Dankook University Hospital) ;
- Yun Hyong Geun (Department of Radiation Oncology Dankook University Hospital) ;
- Jeong Dong Hyeok (Department of Radiation Oncology Wonkwang University Hospital) ;
- Oh Yong Kee (Department of Radiation Oncology Eulji University Hospital) ;
- Kim Jhin Kee (Department of Radiation Oncology Chonbuk National University) ;
- Kim Ki Hwan (Department of Radiation Oncology Chonbuk National University Hospital) ;
- Kim Jeung Kee (Department of Radiation Oncology Dong-A University Hospital)
- 신교철 (단국대학교 의과대학) ;
- 윤형근 (단국대학교 의과대학) ;
- 정동혁 (원광대학교병원 방사선종양학과) ;
- 오영기 (을지대학병원 방사선종양학과) ;
- 김진기 (전북대학교병원 방사선종양학과) ;
- 김기환 (충남대학교병원 방사선종양학과) ;
- 김정기 (동아대학교병원 방사선종양학과)
- Published : 2005.09.01
Abstract
We have designed the multi channel detector for the quality assurance of clinical photon beams. The detector was composed of solid phantom inserted by six plane-parallel ionization chambers at different depth. The chamber as a mini plane parallel chamber was made of carbon coated microfilms. In this study the electrical characteristics of the six chambers in the solid phantom were evaluated using 6 MV photon beam. The leakage currents were less than 0.5 pA, reproducibility was less than 0.5
본 연구에서는 치료용 광자선의 품질관리를 위하여 다채널 측정기를 개발하였다. 측정기는 여러 개의 전리함들이 삽입된 팬텀으로 구성되어 있다. 각 전리함은 탄소가 도포된 마이크로필름으로 제작된 초소형 평행평판 전리함과 같다. 본 연구에서는 고체 팬텀에 삽입된 6개의 전리함에 대하여 6 MV X-선을 이용하여 전기적 특성을 조사하였다 측정결과 누설전류는 0.5 ph 이내로 안정적이었고, 재현성은 0.5