Progress in Medical Physics (한국의학물리학회지:의학물리)
- Volume 16 Issue 2
- /
- Pages.77-81
- /
- 2005
- /
- 2508-4445(pISSN)
- /
- 2508-4453(eISSN)
Characteristics of Detectors for Measurements of Photon Depth Doses in Build-Up Region
선량보강(Build-up) 영역에서의 광자선 깊이선량률 측정을 위한 검출기의 특성 비교
- Kang Sei-Kwon (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Park Suk Won (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Oh Do Hoon (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Park Hee Chul (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Kim Su Ssan (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Bae Hoonsik (Department of Radiation Oncology, College of Medicine, Hallym University) ;
- Cho Byung Chul (Department of Radiation Oncology, College of Medicine, Hallym University)
- 강세권 (한림대학교 의과대학 방사선종양학교실) ;
- 박석원 (한림대학교 의과대학 방사선종양학교실) ;
- 오도훈 (한림대학교 의과대학 방사선종양학교실) ;
- 박희철 (한림대학교 의과대학 방사선종양학교실) ;
- 김수산 (한림대학교 의과대학 방사선종양학교실) ;
- 배훈식 (한림대학교 의과대학 방사선종양학교실) ;
- 조병철 (한림대학교 의과대학 방사선종양학교실)
- Published : 2005.06.01
Abstract
To determine the appropriate method out of various available methods to measure build-up doses, the measurements and comparisons of depth doses of build-up region including the surface dose were executed using the Attix parallel-plate ionization chamber, the Markus chamber, a cylindrical ionization chamber, and a diode detector. Based on the measurements using the Attix chamber, discrepancies of the Markus chamber were within
표면 선량을 포함한 선량보강(build-up) 영역에 대한 깊이선량률(percentage depth dose)을 측정하는 데 있어서 적절한 측정방법을 결정하기 위해 6 MV 광자선에 대해 Atiix와 Markus 평행평판형 이온함, 원통형이온함, 그리고 다이오드 검출기를 사용하여 측정, 비교하었다. Attix 이온함에 의한 측정을 기준으로 할 때, Markus 이온함의 측정은 민조사면(open field)에서
Keywords