Determination of Quality Factors for Cylindrical Ionization Chambers in kV X-rays: Review of IAEA Dosimetry Protocol and Monte Carlo Calculations and Measurements for N23333 and N30001 Chambers
kV X-선에서 원통형전리함의 선질인자 결정에 관한 연구: IAEA 프로토클 고찰과 N23333, N30001 전리함에 대한 몬테칼로 계산 및 측정
- Lee Kang Kyoo (Department of Radiation Oncology, Wonkwang University School of Medicine, Wonkwang Medical Science Research Center) ;
- Lim Chunil (Radiation Standards Division, Korea Food and Drug Administration) ;
- Chang Sei Kyung (Department of Radiation Oncology, College of Medicine, Pochun CHA University Bundand CHA General Hospital) ;
- Moon Sun Rock (Department of Radiation Oncology, Wonkwang University School of Medicine, Wonkwang Medical Science Research Center) ;
- Jeong Dong Hyeok (Department of Physics, Kunsan National University)
- 이강규 (원광대학교 의과대학 방사선종양학교실, 원광의과학연구소) ;
- 임천일 (한국식품의약품안전청 방사선표준과) ;
- 장세경 (포천중문의과대학교 분당차병원 방사선종양학과) ;
- 문성록 (원광대학교 의과대학 방사선종양학교실, 원광의과학연구소) ;
- 정동혁 (군산대학교 물리학과)
- Published : 2005.06.01
Abstract
The quality factors for cylindrical ionization chambers for kV X-rays were determined by Monte Carlo calculation and measurement. In this study, the X-rays of 60-300 kV beam (lSO-4037) installed in KFDA and specified in energy spectra and beam qualities, and the chambers of PTW N23333 and N30001 were investigated. In calculations, the
본 연구에서는 kV X-선에 대하여 원통형전리함의 선질인자를 몬테칼로 계산과 측정으로 결정하였다. 대상 X-선은 KFDA에 설치되어 있으며 에너지스펙트럼과 선질이 명시된