Calibration of High Pressure Vibrating Tube Densimeter

고압 밀도계의 보정에 대한 연구

  • Kim Hwayong (School of Chemical & Biological Engineering, Seoul National University) ;
  • Oh Kyoung Shil (School of Chemical & Biological Engineering, Seoul National University) ;
  • Bae Won (Miwon Commercial)
  • 김화용 (서울대학교 화학생물공학부) ;
  • 오경실 (서울대학교 화학생물공학부) ;
  • 배원 (미원상사주식회사)
  • Published : 2005.09.01

Abstract

This work was to calibrate Anton Paar 512P vibrating tube densimeter which is a special type for density measurement under high pressure (up to 70 MPa) and temperature (up to 420 K) and checked the validity of the new apparatus. Calibration was peformed with water, nitrogen and water, vacuum as reference fluids. Temperature is measured from 313.15 K and 353.15 K at 10 K intervals, and pressure is measured up to 30 MPa at 5 MPa intervals, respectively. To check the validity of new apparatus and the calibration method we compared the values obtained by measuring density of carbon dioxide with those given by NIST Thermophysical Properties of Pure Fluids Database. The average deviations between the NIST and the experimental data were $0.3\%$ with water and nitrogen as reference fluids.

본 연구에서는 넓은 범위의 온도(420 K), 압력(70 MPa)에서 밀도를 측정할 수 있는 Anton Paar DMA 512P vibrating tube densimeter를 보정하고, 장치의 건전성을 확인하였다. 보정을 위해서 기준물질로 물과 질소 그리고 물과 진공의 period를 측정하여 각각을 비교해보았다. 온도는 313.15부터 353.15 K까지 10 K 간격으로 5 isotherms를 측정하였고, 압력은 5부터 30 MPa까지 대략 5 MPa 간격으로 측정하였다. 또한 이산화탄소의 밀도를 측정하여 NIST에 보고된 밀도값과 비교해 보았다. 기준물질로 물과 진공보다 물과 질소를 사용하였을 때 더 좋은 결과를 보였고, 이산화탄소의 경우 측정범위 내에서는 오차가 대략 $0.3\%$로 비교적 낮은 결과를 보였으며 이로써 장치의 건전성을 확인하였다.