A study on the utilization of CT equipments

전산화 단층촬영 장치의 이용 실태에 대한 조사

  • Cho, Pyong-Kon (Dept. of Diagnostic Radiology, Korea University Medical Center) ;
  • Oh, Yu-Whan (Dept. of Diagnostic Radiology, Korea University Medical Center) ;
  • Kim, Sung-Soo (Dept. of Radiologic Technology, College of Singu) ;
  • Choi, Jong-Hak (Dept. of Radiologic Technology, College of Health Sciences, Korea University) ;
  • Kim, You-Hyun (Dept. of Radiologic Technology, College of Health Sciences, Korea University)
  • 조평곤 (고려대학교 안암병원 진단방사선과) ;
  • 오유환 (고려대학교 안암병원 진단방사선과) ;
  • 김성수 (신구대학 방사선과) ;
  • 최종학 (고려대학교 병설 보건대학 방사선과) ;
  • 김유현 (고려대학교 병설 보건대학 방사선과)
  • Published : 2005.03.30

Abstract

This study was conducted to investigate the patient exposure dose during a CT examination and the present position in utilization of the CT equipments. To this end Questionnaire were sent out to 278 medical facilities registered at Korea Hospital Association and replies of 161 medical facilities were analyzed. The results were as follows ; 1) The distribution of CT examination was. Brain 40.7%, Abdomen 27.8%, Chest 15.7%, Spine 4.8%, Pelvis 4.1%, PNS 3.2%, Facial bone 2.5% and Extremity 1.1% respectively. 2) The statistics for dose index according to each part of examination were Brain 38.0mGy, Abdomen 12.0mGy for adults : Brain 13.6mGy, Abdomen 6.2mGy for infant. Our surveyed dose index appeared lower than the IAEA recommends. 3) Most medical facilities have selected the parameters for radiographic exposure in the range of $100{\sim}120kVp,\;100{\sim}250mA\;and\;1{\sim}2$ seconds.

전산화 단층촬영검사 시 환자가 받는 피폭선량과 장치 이용 실태를 파악하기 위한 목적으로 대한병원협회에 등록된 전국의 278개 의료기관에 설문지를 발송하여 설문에 응답한 161개 의료기관의 자료를 분석하여 다음과 같은 결과를 얻었다. 1. 검사의 부위별 촬영분포는 Brain(40.7%), Abdomen(27.8%), Chest(15.7%), Spine(4.8%), Pelvis(4.1%), PNS(3.2%) Facial bone(2.5%), Extremity(1.1%)등으로 Brain검사가 가장 많았다. 2. 검사 부위별 선량지수에 대한 통계는 성인의 경우 Brain 38.0mGy, Chest 10.1mGy, Abdomen 12.0mGy, Pelvis 13.2mGy, PNS 27.5mGy, Facial bone 28.6mGy, Spine 26.0mGy, Extremity 13.5mGy이었고, 유아의 경우 Brain 13.6mGy, Abdomen 6.2mGy 이었다. 3. 검사부위별 촬영조건에 대한 통계는 관전압 $100kVp{\sim}120kVp$, 관전류 $100mA{\sim}250mA$, 조사시간은 1초${\sim}$2초를 가장 많이 사용하고 있었다.

Keywords

References

  1. National Institute of Radiological Science (supervisr), Report 1998 to the general Meeting of the International Science Committee on the Source, Effect and Risk of Radiation-Effect of Atomic Radiation Attachment C(exposure from the medical use of radiation). Jitsugyokoho, Tokyo. 265-334(1990)
  2. IAEA, International basic sagety standards for protection against ionizing radiation and the safety of radiation source. 279-280, EADA Safety Series No.115, Vienna(1996)
  3. T. Orito, Investigation of accrual situation of radiological techniques to take measures for patient exposure dose reduction. Jpn. J. Radiological technology 32, 34-39(1974)
  4. T. Orito, S. Sanada, S. Koga, et al., Changes in x-ray diagnostic technology and patient exposure dose comparison with 1973 and 1979. A Monthly Journal of Medical Imaging and Information 12(6), 325-327(1980)
  5. S. Suzuki, T. Orito, Koga, et al., Changes in x-ray diagnostic technology and patient exposure dose comparison with 1973 and 1979. A Monthly Journal of Medical Imaging and Information 22(6), 359-363(1990)
  6. 조평곤, 전산화 단층촬영장치의 정도관리에 관한 연구 석사학위논문 고려대학교 교육대학원(2001)
  7. J. Anthony Seibert, Gray T. Barnes, Robert G. Gould, et al., Specification, Acceptance testing and quality control of diagnostic X-ray imaging equipment. Medical physics monograph No. 20, 899-936(1991)
  8. Robert L. Mini, et al., Radiation exposure of patiants who undergo CT of the trunk. Radiology Volume 195 No.2, 557-562(1995)
  9. D. J. Brenner, C. D. EUiston, E. J. Hall, et al, Estimate risks of radiation induced fatal cancer from pediatric CT. AJR 176, 289-296(2001) https://doi.org/10.2214/ajr.176.2.1760289
  10. 최종학 외, 의료방상선 생물학, 신광출판사(2004)
  11. G. Bongartz, S. J. Golding, A. G. Jurik, M. Leonardi, E. v. P. van Meerten, J. Geleijns, K A. Jessen, W. Panzer, P. C. Shrimpton, G. Tosi et al., European guidelines on quality criteria for computed tomography. EUR 16262 EN
  12. 'Operating Instructions Somatom Plus CT, Dose and Imaging Performance Information'. Siemens Aktiengesellschaft(1990)
  13. T. B. Shope, R. M. Gange, and G. C. Johnson, 'A method for describing the doses delivered by transmission x-ray computed tomography', Med. Phys.8, 488(1981)
  14. T. B. Shope, T. J. Morgan, C. K Showalter, K. S. Pentiow, L. N. Rothenberg, D. R White, and R D. Speller, 'Radiation dosimetry survey of computed tomography systems from ten manufacturers', Br. J. Radiol. 55, 60(1982)
  15. IAEA Guidance level(2002)
  16. K.S. Pentiow, 'Dosimetry in computed tomography', Chapter 124, Volume Five Radiology of the Skull and Brain, Technical Aspect of Computed Tomography, T.H. Newton and D.G. Potts, Eds., C.V. Mosby, St. Louis(1981)