대한임상검사과학회지 (Korean Journal of Clinical Laboratory Science)
- 제36권2호
- /
- Pages.98-109
- /
- 2004
- /
- 1738-3544(pISSN)
- /
- 2288-1662(eISSN)
자동화학 분석기 Toshiba 120 FR의 평가
Evaluation of Autochemical Analyzer Toshiba 120 FR
- 박점기 (대전보훈병원 진단검사의학과)
- Park, Jum Gi (Department of Laboratory Medicine, Daejeon Veterans Hospital)
- 발행 : 2004.12.31
초록
The lower limit of detection, precision, setting method of target value, reportable range determination, recovery, linearity, and comparison study with another equipment was evaluated for the Toshiba-120FR chemistry autoanalyzer which was newly introduced at the Daejeon Veteran Hospital in Dec. 2003. Nineteen kinds of test for AST, ALT, ALP, LDH, GGT, TP, ALB, GLU, T-cho, T-bil, TG, UA, CAL, IP, AMY, HDL-C, LDL-C, Cre and BUN were performed to evaluate the lower limit of detection, precision, setting method of target value, reportable range determination, recovery, linearity, and comparison study with other equipment according to the NCCLS guidelines(EP5-A, EP6-P, EP9-A). The Toshiba-120FR autochemical analyzer showed good precision for all tested items. The data concerning the lower limit of detection, precision(total CV 0.47%~3.65%), setting method of target value, reportable range determination, recovery(93%~111%), linearity(