Clinical Implementation of a Virtual-Micro MLC for Smoothing MLC Field Edge

다엽콜리메이터에 의한 조사경계면에서의 요동현상을 완화시켜주는 가상미세다엽콜리메이터의 임상 적용

  • 조병철 (한림대학교 의과대학 방사선종양학교실, 한림대학교 성심병원 방사선종양학과) ;
  • 박희철 (한림대학교 의과대학 방사선종양학교실, 한림대학교 성심병원 방사선종양학과) ;
  • 배훈식 (한림대학교 의과대학 방사선종양학교실, 한림대학교 성심병원 방사선종양학과)
  • Published : 2004.09.01

Abstract

A Siemens HD-270 MLC$^{TM}$, a virtual-micro MLC, allows to the application of a smooth field edge method due to the finite leaf-width of MLC. This technique was implemented into a Pinnacle planning system in order to evaluate the dose distributions during the planning stage. The necessary dosimetric aspects, such as undulation and effective penumbra, were investigated with variations in the resolutions of a virtual-micro MLC and field edge angle. The positional accuracy of the couch movement was also assessed for clinical implementation. The overhead time for planning and treatment was confirmed as negligible.e.

지멘스(Siemens)사에서 개발된 가상미세다엽콜리메이터인 HD-270은, 전형적 다엽콜리메이터에 있어 폭의 유한성 때문에 불필요하게 발생하는 조사경계면의 요동현상을 완화시켜준다. 저자들은 이 기법을 사용했을 때의 선량분포의 변화를 확인하고, 치료계획단계에서 이를 평가해 볼 수 있도록 Pinnacle (Philips Radiation Oncology Systems, 미국) 전산화치료계획장치에 구현하였다. 그리고 임상적용에 앞서 선량학적 특성을 파악하기 위해, 가상미세다엽콜리메이터의 해상도, 조사경계면과 다엽콜리메이터가 형성하는 각도가 요동현상 및 유효반음영에 미치는 영향을 평가하였다. 또한 가상미세다엽콜리메이터 구현에 수반되는 치료테이블 움직임에 대한 위치 정확도를 평가하였다. HD-270의 임상적용을 위해 추가적으로 소요되는 치료계획과 치료 시간은 무시할 수 있음을 확인할 수 있었다.