The Isoflurane Concentration of Precision Vaporizer Goldman Vaporizer According to Room Temperature and Carrier Gas Flow Rate

온도 및 유량에 따른 흡입 마취제 전용 기화기와 Goldman 기화기의 isoflurane 농도 변화

  • 김성미 (건국대학교 수의과대학 수의외과학교실) ;
  • 장화석 (건국대학교 수의과대학 수의외과학교실) ;
  • 이정선 (건국대학교 수의과대학 수의외과학교실) ;
  • 최치봉 (건국대학교 수의과대학 수의외과학교실) ;
  • 임희란 (건국대학교 수의과대학 수의외과학교실) ;
  • 최준철 (건국대학교 수의과대학 수의외과학교실) ;
  • 김휘율 (건국대학교 수의과대학 수의외과학교실)
  • Published : 2004.09.01

Abstract

Anesthetic machines may be equipped with either a precision or nonprecision vaporizer. A precision vaporizer is designed to deliver an exact concentration of anesthetic agent. Goldman vaporizer is a low-flow, closed-circle circuit with a low resistance vaporizer, in circuit. Vaporizers used within circle system(VIC) are not usually temperature compensated and this is generally thought to be a disadvantage. As the volatile agent is vaporized, heat is extracted from the liquid and temperature decreases. This cooling of the liquid leads to a decrease in concentration of the anaesthetic agent delivered by the vaporizer. The purpose of this study is to examine the mechanical consistency of the delivery of isoflurane from Goldman vaporizer and precision vaporizer at various gas flow rates and temperatures. And we first studied isoflurane concentration according to room temperature changes delivered by a Goldman vaporizer and precision vaporizer using different gas flow. The room temperature of $15^{\circ}C,$ $20^{\circ}C,$ $28^{\circ}C$ and fresh gas flow rates of 0.5, 1.0, 1.5, 2.0, 3.0 l/min were used. The inspired agent concentration was measured using a Datex-Ohmeda multigas analyzer. As rose in room temperature, the isoflurane concentration of precision vaporizer approximated the dial setting. On the other hand, at a dial setting concentration of 5.0 percent the delivered isoflurane concentration of precision vaporizer was more than the dial setting in high temperature. The isoflurane concentration of precision vaporizer remained constant despite the increase in temperature. The isoflurane concentration of Goldman vaporizer was increased with rise in room temperature and decreased with rise in gas flow.

Keywords

References

  1. Andrews JJ, Johnston RV Jr, Kramer GC. Consequences of misfilling contemporary vaporizers with desflurane. Can J Anaesth 1993; 40(1): 71-76 https://doi.org/10.1007/BF03009323
  2. Bednarski RM, Gaynor JS, Muir WW. 3rd. Vaporizer in circle for delivery of isoflurane to dogs. J Am Vet Med Assoc 1993; 202(6): 943-948
  3. Bewley JS, Eltringham RJ, Sanderson P. Closed circuit anaesthesia in ventilated patients using the Komesaroff vaporizer within the circle. Anaesth Intensive Care 1998; 26(5): 558-562
  4. Brosnan S, Royston B, White D. Isoflurane concentrations using uncompensated vaporizers within circle systems. Anaestehsia 1998; 53(6): 560-564 https://doi.org/10.1046/j.1365-2044.1998.00366.x
  5. Craig GR, Berry CB, Yeats MJ. An evaluation of the Universal PAC and oxford miniature vaporizers for paediatric field anaesthesia. Anaesthesia 1995; 50(9): 789-793 https://doi.org/10.1111/j.1365-2044.1995.tb06142.x
  6. Feldman SA, Morris LE. Vaporization of halothane and ether in the copper kettle. Anesthesiology 1958; 19(5): 650-655 https://doi.org/10.1097/00000542-195809000-00008
  7. Garden AL, Haberkern CM, Buckon ME, Baptiste DV. Output from a Drager vapor 19.1 on a moving trolley. 1997; 44(2): 227-228
  8. Gusterson FR, Clark JM. The use of the Goldman halothane vaporizer in the closed circuit. Br J Anaesth 1960; 32: 388-393 https://doi.org/10.1093/bja/32.8.388
  9. Harrison GR. The contamination of volatile anaesthetics in an in-circle vaporizer with water during prolonged closed-circle anaesthesia. Anaesthesia 2000; 55(8): 791-792 https://doi.org/10.1046/j.1365-2044.2000.01131.x
  10. Komesaroff D. A new anesthetic machine and technic with particular application to developing areas. Anesth Analg 1973; (4): 605-613
  11. Lin CY. Assessment of vaporizer performance in low-now and closed-circuit anesthesia. Anesth Analg 1980; 59(5): 359-366
  12. McKelvey D, Hollingshead KW. Anesthetic equipment. In: Small animal anesthesia and analgesia, 2nd ed. St.Louis: Mosby. 2000: 147-190
  13. Morris LE. A new vaporizer for liquid anesthetic agents. Anesthesiology 1952; 13(6): 587-593 https://doi.org/10.1097/00000542-195211000-00003
  14. Muir WW, Hubbell JAE, Skarda RT, Bednarski RM. Anesthetic machines and breathing systems. In: Handbook of veterinary anesthesia, 3rd ed. St. Louis: Mosby. 2000: 210-231
  15. Paddleford RR. Anesthetic equipment. In: Manual of small animal anesthesia, 2th ed. New York: Churchill Livingstone. 2000: 89-109
  16. Piedra AG, Pereira JL, Agut A. Efficacy of the Komesaroff anaesthetic machine for delivering isonurane to dogs. Vet Rec 1998; 143(16): 437-440 https://doi.org/10.1136/vr.143.16.437
  17. Revell SP, Tayler DH. Isoflurane in a circle system with low gas flow. Br J Anaesth 1987; 59(10): 1219-1222 https://doi.org/10.1093/bja/59.10.1219
  18. Richardson W, Carter KB. Evaluation of keyed fillers on TEC vaporizers. 1986; 58(3): 353-356 https://doi.org/10.1093/bja/58.3.353
  19. Robins DW. Closed system enflurane in oxygen. Anaesthesia 1983; 38(1): 56-60 https://doi.org/10.1111/j.1365-2044.1983.tb10375.x
  20. Rodgers RC, Hill GE. Equation for vapour pressure versus temperatures ; derivation and use of the Antoine equation on a hand-held programmable calculator. British Journal of Anaesthesia 1978; 50: 415-424 https://doi.org/10.1093/bja/50.5.415
  21. Stoelting RK, Miller RD. Anesthesia systems. In: Basics of Anesthesia, 3rd ed. New York: Churchill Livingstone. 1994: 131-147
  22. Stokes JW. Use of air to maintain nitrogen stores during low flow and closed circuit anaesthesia. Anaesth Intensive Care 1994; 22(4): 391-393
  23. Thurmon JC, Tranquilli WJ, Benson GJ. Equipment and monitoring. In: Essentials of small animal anesthesia and analgesia. Philadelphia: Wolters Kluwer. 1999: 225-291
  24. Waters RM. Clinical scope and utility of carbon dioxide infiltration in inhalation anesthesia. Anesth Analg 1924; 3: 20
  25. Weir HM, Kennedy RR. Infusing liquid anaesthetics into a closed circuit. Anaesth Intensive Care 1994; 22(4): 376-379
  26. White SA, Strunin L. Anaesthesia with the Boyle's bottle vaporizer. Anaesthesia 1996; 51(10): 939-942 https://doi.org/10.1111/j.1365-2044.1996.tb14962.x
  27. William W, Muir III, Jennifer Gadawski. Cardiorespiratory effects of low-flow and closed circuit inhalation anesthesia, using sevoflurane delivered with an in- circuit vaporizer and concentrations of compound A. Am J Vet Res 1998; 59(5): 603-608
  28. Wright D, Bronsnan S, Royston B, White D. Controlled ventilation using isoflurane with an in-circle vaporizer. Anaesthesia 1998; 53(7): 650-653 https://doi.org/10.1046/j.1365-2044.1998.365-az0510.x
  29. Young DA, Brosnan SG, White DC. A semiquantitative analysis of the dynamics of a Glodman-type vaporizer. 2000; 55(6): 557-570 https://doi.org/10.1046/j.1365-2044.2000.01318.x
  30. 김수연, 이윤우, 이영석, 김진수, 남승철, 김종례. 불량 기화기로 인한 할로탄의 과량투여. 대한마취과학회지 1988; 21: 1019-1023
  31. 김순임. 할로텐 흡입농도에 대한 임상적 연구. 대한마취과학회지 1984; 17: 27-32
  32. 대한마취과학회 교과서 편집위원회. 마취과학. 서울: 여문각., 2002: 46-57
  33. 배기열, 류길수, 김향미, 주진철. 개인 병원 마취기 기화기의 정확도에 관한 조사. 대한마취과학회지 1996; 30: 658-662
  34. 승익상. PhysioFlex 마취기를 이용한 폐쇄식과 반폐쇄식 마취방법에 따른 마취제 소모량 비교. 대한마취과학회지 1998; 34: 758-763
  35. 승익상. 폐쇄식 마취용 physioFlex 마취기의 임상 사용경험. 대한마취과학회지 1999; 36: 571-576