Abstract
The purpose of this study is to develop an ozone passive sampler and to evaluate its performance p-Acetamidophenol using as the reagent for ozone reacts specifically with the ambient ozone to produce a fluorescence material (p-acetamidopheonl dimer). The volume of absorbent solution and the extraction time determined at suitable conditions for measuring ozone were 100$\mu$L and 60 min, respectively. The changes of fluorescence were observed with incresing the storage period of passive samplers in ambient air. but the cool storage in a refrigerator did not remarkably influence the increase of fluorescence. The measurement for the precision oi the passive sampling was carried out with duplicate measurement of passive samplers. The intra-class correlation coefficients of passive samplers using dry and wet filters were 0.992 and 0.962, respectively The results from field validation tests indicated practical agreement (dry filter: r=0.963, wet filter: r=0.995) between the passive sampler and an UV photometric $O_3$ analyzer. The limit of quantification of ozone passive samplers with sampling time of 8 hr (wet filter) and 24 hr (dry filter) were 8.0 ppb and 2.7 ppb, respectively.