Usefulness of inverter type REX-650 R/F X-ray equipments

REX-650 RF 인버터장치의 성능평가

  • Kim, Jung-Min (Dept. of Radiological technology, College of health Science, Korea University) ;
  • Oh, Jung-Hwan (Dept. of Radiological technology, Gachongil College)
  • 김정민 (고려대학교 보건대학 방사선과) ;
  • 오정환 (가천길대학 방사선과)
  • Published : 2003.03.30

Abstract

The conclusions after appraisal usefulness of korean 30 kHz inverter type X-ray equipments which are currently used in the clinics are like those followings. 1. The specific characters of mA, kVp, msec in the radiography keep very good accuracy and showed good waveform also under KS regulations. 2. The output of the X-ray equipment is showed 1.7 times higher than single-phase. 3. mA, kVp of the fluoroscopy and Sopt film exposure mA, kVp showed appropriated results under the regulations. 4. As we consider the conversion factor of image intensifier tube used for two years, it showed pretty high results, 310 at 75 kV, and knew that the conversion factor which followed by increasing kVp increased its computations like Kim's experimentation. Those X-ray equipment showed excellent results in the appraisal while those were operating so we think that those X-ray equipments will be substitute for expensive foreign equipment in korean domestic medical equipment in the future.

현재 임상에서 사용중인 국산 30 kHz 용량의 PWM방식의 인버터식 X선 장치의 성능을 장치의 출력정확도, 출력파형의 변화 및 영상증배관의 변환계수 등을 평가하여 그 유용성을 알아보았다. 그 결과 촬영용 장치의 관전류, 관전압, 단시간 특성은 KS규정에 대해 매우 양호한 정확도를 유지했고 출력파형도 양호한 특성을 나타내었다. 또한 촬영시 X선 장치의 출력은 단상대비 약 1.7배 정도 많게 나타났다. 투시장치의 성능에서도 관전류, 관전압 및 sopt촬영 관전류, 관전압 모두 최고 5.8% 이하의 양호한 결과를 나타내었으며, 영상증배관의 변환계수는 2년여 사용한 것을 고려할 때 75kV에서 310으로 상당히 높았다. 국내기술로 개발된 고주파식 X선 장치가 성능의 실용적 단계를 넘는 것으로 나타나 앞으로 고주파식 X선 장치가 국내, 외에 많이 보급되고 이용될 것으로 사료된다.