초음파 가열 시 In Vitro 및 In Vivo에서 Microwave Radio-Thermometer와 탐침온도계의 일치도
In Vitro and In Vivo Agreement of Microwave Radio-Thermometer and Needle Probe Thermometer During Therapeutic Ultrasound
- 이수영 (연세대학교 대학원 재활학과) ;
- 조상현 (연세대학교 보건과학대학 물리치료학과 및 보건과학연구소) ;
- 이충휘 (연세대학교 보건과학대학 물리치료학과 및 보건과학연구소) ;
- 김종만 (서남대학교 물리치료학과)
- Lee, Su-Young (Dept. of Rehabilitation Therapy, The Graduate School, Yonsei University) ;
- Cho, Sang-Hyun (Dept. of Physical Therapy, College of Health Science, Yonsei University, Institute of Health Science, Yonsei University) ;
- Yi, Chung-Hwi (Dept. of Physical Therapy, College of Health Science, Yonsei University, Institute of Health Science, Yonsei University) ;
- Kim, Jong-Man (Dept. of Physical Therapy, Division of Health, Seonam University)
- 발행 : 2003.02.19
초록
Therapeutic ultrasound is commonly applied for deep heating in physical therapy setting. However, it is difficult to determine the exact application dosage and to confirm the immediate heating effect. Microwave Radio-Thermometer (MRT) can measure the temperature by the electromagnetic energy in the microwave region of the object that emits above absolute zero temperature. MRT was used for early diagnosis of breast cancer since it was not harmful, non-invasive, and non-ionizing to the human body. The purposes of this study were to investigate how accurately 1.1 GHz RTM (RES Ltd. Russia) measures the change of average temperature in the tissue, and to determine the depth of temperature change measurement. Therapeutic ultrasound was applied (continuous wave for 5 minutes, 1 MHz, intensity of 1.5