Calibration and a Plane of Incidence Fixed Ellipsometer

Ellipsometry에서의 Calibration 및 입사면 고정형 Ellipsometer

  • Published : 2003.12.01

Abstract

The general users find difficulties in using ellipsometers. Thus, the object of this study is to construct an ellipsometer with simple operation principle. We developed an ellipsometer which does not require alignment and calibration before measuring sample. A basis structure model after rotating a compensator spectroscopic ellipsometry, the fixed incidence angle at $70^{\circ}$. This ellipsometer does not demand calibration, because the plane of incidence is not changed due to the novel sample holder structure. The results for various standard samples were compared with those from conventional RCSE to test the performance of this instrument. Also repeated measurements were performed to test the precision of the calibration coefficient in a plane of incidence fixed.

Keywords