Abstract
A transformation methode of the chromaticity coordinates was proposed to calibrate the measured data obtained by a urine analysis system which implemented in our previous study. Generally. the reacted color of a reagent strip by urine analysis system often exhibit the color distortions due to nonlinear characteristics of the various devices that is the optic module mechanism. hardware, and surround circumstance. A color correction method for minimizing the color distortion play a few role in maintaining high accuracy and reproduction of the urine analysis system. In this work, we used the compensation method such as the shading correction, the characteristic curve extraction of RGB color by means of third order spline interpolation, and linear transformation using a reference color. In addition, 1931 CIE XYZ color space was used to compensate the color of the measured data by a standard reference system as colorimeter. A compensation matrix was obtained so that the output values of the urine analysis system is nearly equal to that of a standard reference system for identical color sample. Color correction obtained by a urine analysis system which implemented in our previous study exhibited a good color accuracy when it was compared with the reference data. Observed result from an experiments on ten items or a urinalysis strip that color difference or between two urine analysis system was 1.28.
본 연구에서는 구현된 요분석 시스템으로 획득한 데이터를 보정하기 위하여 색 좌표 변환 기법을 제안하였다. 일반적으로 요분석 시스템은 요분석용 스트립의 정색반응을 검출하는 과정에서 여러 가지 비선형적인 특성 즉 광 모듈의 메커니즘, 하드웨어, 그리고 주변 환경에 의해 색 왜곡된 입·출력 특성을 지닌다 따라서 보다 높은 정확도와 재현성을 유지하기 위해 장비 특성화 기법을 도입하여 색 왜곡 현상을 보정하였다. 본 연구에서는 명암 보정, ,3차 스플라인 보간법에 의한 RGB 신호의 특성 곡선 추출, 기준색 고정 선형변환 기법을 사용하여 색 보정 과정을 수행하였다 색 보정을 위해 사용된 표준 장비는 1931년 CIE XYZ 색공간 특성을 지닌 좌표계로 설정하였으며, 동일한 칼라 샘플에 대해 요분석 시스템의 출력값과 표준 장비의 출력값이 일치되도록 하는 보정 행렬을 구하였다. 구현된 요분석 시스템을 색 보정한 후 기준 데이터와 비교한 결과 양호한 색상 정확도를 나타내었다 요분석용 스트립의 10 가지 항목에 대해 구현된 두 대의 요분석 시스템을 사용하여 실험한 결과 장비간 색차는 1.28이었다.