The characteristics on dose distribution of a large field

넓은 광자선 조사면($40{\times}40cm^2$ 이상)의 선량분포 특성

  • Lee Sang Rok (Dept. of Radiation Oncology, Anam Hospital, Korea University Medical Center) ;
  • Jeong Deok Yang (Dept. of Radiation Oncology, Anam Hospital, Korea University Medical Center) ;
  • Lee Byoung Koo (Dept. of Radiation Oncology, Anam Hospital, Korea University Medical Center) ;
  • Kwon Young Ho (Dept. of Radiation Oncology, Anam Hospital, Korea University Medical Center)
  • 이상록 (고려대학교 의료원 안암병원 방사선종양학과) ;
  • 정덕양 (고려대학교 의료원 안암병원 방사선종양학과) ;
  • 이병구 (고려대학교 의료원 안암병원 방사선종양학과) ;
  • 권영호 (고려대학교 의료원 안암병원 방사선종양학과)
  • Published : 2003.09.01

Abstract

I. Purpose In special cases of Total Body Irradiation(TBI), Half Body Irradiation(HBI), Non-Hodgkin's lymphoma, E-Wing's sarcoma, lymphosarcoma and neuroblastoma a large field can be used clinically. The dose distribution of a large field can use the measurement result which gets from dose distribution of a small field (standard SSD 100cm, size of field under $40{\times}40cm2$) in the substitution which always measures in practice and it will be able to calibrate. With only the method of simple calculation, it is difficult to know the dose and its uniformity of actual body region by various factor of scatter radiation. II. Method & Materials In this study, using Multidata Water Phantom from standard SSD 100cm according to the size change of field, it measures the basic parameter (PDD,TMR,Output,Sc,Sp) From SSD 180cm (phantom is to the bottom vertically) according to increasing of a field, it measures a basic parameter. From SSD 350cm (phantom is to the surface of a wall, using small water phantom. which includes mylar capable of horizontal beam's measurement) it measured with the same method and compared with each other. III. Results & Conclusion In comparison with the standard dose data, parameter which measures between SSD 180cm and 350cm, it turned out there was little difference. The error range is not up to extent of the experimental error. In order to get the accurate data, it dose measures from anthropomorphous phantom or for this objective the dose measurement which is the possibility of getting the absolute value which uses the unlimited phantom that is devised especially is demanded. Additionally, it needs to consider ionization chamber use of small volume and stem effect of cable by a large field.

I. 목적 Total Body Irradiation(TBI)와 Half Body Irradiation(HBI), Non-Hodgkin's lymphoma, E-Wing's sarcoma, lymphosarcoma, neuroblastoma 등의 특수한 경우에 넓은 광자선 조사면($40{\times}40cm2$ 이상)이 임상적으로 사용될 수 있다. 넓은 광자선 조사면의 선량분포는 매번 실제 측정 대신 좁은 광자선 조사면 (표준 SSD 100cm, 조사면의 크기 $40{\times}40cm2$ 미만)에서 얻은 측정결과를 이용하여 보정할 수 있으나, 단 순한 계산에 의한 방법만으로는 산란 방사선의 여러 가지 요인에 의한 실제 신체 각 부위의 선량 및 그 균일성을 알기는 힘들다. 본 연구에서는 치료거리 증가에 따른 넓은 광자선 조사면의 기본 parameter(PDD, TMR, Output, Sc, Sp)를 측정하고, 좁은 광자선 조사면에서 얻은 측정결과와 비교하여 그 차이를 확인해 보고 실제 적용여부를 알아보고자 한다. II. 대상 및 방법 표준 SSD 100cm에서 Multidata water phantom을 이용하여 조사면의 크기 변화에 따라 기본parameter(PDD, TMR, Output, Sc, Sp)를 측정하였다. 먼저 SSD 180cm에서(phantom이 치료실 바닥으로 수직방향) 조사면 증가에 따른 기본 parameter를 측정하였고, SSD 350cm에서(phantom이 치료실 벽면으로 수평방향이고 horizotal beam의 측정이 가능한 mylar를 가진 small water phantom을 이용) 같은 방법으로 측정하여 서로 비교해보았다. III. 결과 및 결론 SSD 180cm과 350cm에서 측정한 parameter들이 표준 선량측정 자료와 비교해서 오차범위가 실험적 오차에 있을 정도로 큰 차이가 없었음을 알 수 있었다. 정확한 자료를 얻기 위해 anthropomorphous phantom에서 선량측정을 하거나 이런 목적을 위해 특별히 고안된 unlimited phantom을 이용한 절대값을 얻을 수 있는 선량 측정이 요구된다. 부가적으로 작은 부피의 ionization chamber 사용과 넓은 조사면에 의한 cable과 stem effect를 고려해야 할 필요가 있다.