Tuberculosis and Respiratory Diseases
- Volume 53 Issue 4
- /
- Pages.431-438
- /
- 2002
- /
- 1738-3536(pISSN)
- /
- 2005-6184(eISSN)
The Usefulness and Safety of Natural Stent in a Canine Model of Tracheal Stenosis
Nd-YAG laser를 이용한 기관협착 동물모델에서 Natural 스텐트의 안전성 및 유효성 -Dumon 스텐트와의 비교 실험-
- Kim, Ho-Joong (Division of Pulmonary Critical Care Medicine Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
- Koh, Won-Jung (Division of Pulmonary Critical Care Medicine Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
- Suh, Gee-Young (Division of Pulmonary Critical Care Medicine Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
- Chung, Man-Pyo (Division of Pulmonary Critical Care Medicine Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
- Kim, Jhin-Gook (Department of Thoracic Surgery, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
- Suh, Soo-Won (Department of Medical Engineering, Samsung Medical Center, Sungkyunkwan University School of Medicine) ;
- Kwon, O-Jung (Division of Pulmonary Critical Care Medicine Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine)
- 김호중 (성균관대학교 의과대학 삼성서울병원 호흡기내과학교실) ;
- 고원중 (성균관대학교 의과대학 삼성서울병원 호흡기내과학교실) ;
- 서지영 (성균관대학교 의과대학 삼성서울병원 호흡기내과학교실) ;
- 정만표 (성균관대학교 의과대학 삼성서울병원 호흡기내과학교실) ;
- 김진국 (성균관대학교 의과대학 삼성서울병원 흉부외과학교실) ;
- 서수원 (성균관대학교 의과대학 삼성서울병원 의공학과학교실) ;
- 권오정 (성균관대학교 의과대학 삼성서울병원 호흡기내과학교실)
- Published : 2002.10.30
Abstract
Background : In order to investigate the usefulness and safety of Natural Stent, we performed this study in a canine model of tracheal stenosis induced using Nd-YAG laser. Materials and Methods : After tracheal stenosis was induced in 12 Mongrel dogs using Nd-YAG laser, either Dumon (n=6) or Natural (n=6) stent was inserted into the trachea. To assess the degree of stent migration and mucostasis, bronchoscopy was performed every week for 4 weeks, after which all stents were removed. One week after stent removal, tracheal stenosis was evaluated by bronchoscopy. Results : The degree of stent migration was not different between the dogs with Dumon stent (
배 경 :기체역학적 실험을 기반으로 새로 개발된 Natural 스텐트(N-Stent)의 유효성과 안전성을 기존의 Dumon 스텐트와 비교하고자, Nd-YAG Laser를 이용한 한국산 잡견 기관협착 동물모델에서 연구를 시행하였다. 방 법 : 한국산 잡견 12마리를 Nd-YAG 레이저를 이용하여 기관협착을 유도한 후, 무작위로 선정한 6마리에게는 Dumon 스텐트를, 6마리에게는 Natural 스텐트를 삽입하였다. 실험동물은 매 1주마다 4주 동안 기관지내시경으로 스텐트 위치변동과 점액저류 정도를 반정량적으로 기록하였고, 삽입 4주후에 스텐트를 제거하였다. 스텐트 제거 1주 후, 기관협착 정도를 조사하였다. 결 과 : Natural 스텐트를 삽입한 동물에서 스텐트의 위치 변동의 정도(