A Study on the Image Quality of Mammography and the Average Glandular Dose
맘모그래피의 화질과 평균유선조직선량에 관한 검토
- Lee, In-Ja (Dept. of Radiological Technology Dong Nam College) ;
- Kim, Hak-Sung (Dept. of Radiological Technology Dong Nam College) ;
- Kim, Sung-Soo (Dept. of Radiological Technology Shin Gu College) ;
- Huh, Joon (Institute of Radiological Technology Listem Co., LTD)
- Published : 2002.12.31
Abstract
We came to the following conclusion as the results of experiment on the image quality of mammography and the average glandular dose using 4 apparatuses at 3 hospitals in Seoul. 1. Whereas the measurement of half value layer showed no differences among the apparatuses, the measurement by an attenuation curve method showed some differences by 5.9%. There were 9.1% differences in the measurement by aluminum conversion method. 2. The basic density of an automatic exposure control unit must be D = 1.40, but there was no automatic exposure unit adjusted precisely at any hospitals. The unit at the B hospital exceeded the allowable limit by
유방촬영에 대한 화질과 평균유선조직선량에 대하여 실험한 결과 반가층은 장치마다 큰 차이를 나타내지 않았으며, 자동노출기구를 사용할 경우 정확성에 문제가 있으므로 확인이 필요하다. 유방촬영에 사용되는 자동현상기 또한 꼭 관리가 필요하며, 유방촬영 시 사용되는 증감지/필름에 따라서 팬톰 촬영 식별능에 많은 영향을 받고 있으므로 선택 시 신중을 기해야 한다. 유방두께 4.2 cm에 대한 평균유선조직 선량은 관전압에 따라 달라지며, ACR기준인 3.0 mGy보다 훨씬 적은