A Study on the Image Quality of Mammography and the Average Glandular Dose

맘모그래피의 화질과 평균유선조직선량에 관한 검토

  • Lee, In-Ja (Dept. of Radiological Technology Dong Nam College) ;
  • Kim, Hak-Sung (Dept. of Radiological Technology Dong Nam College) ;
  • Kim, Sung-Soo (Dept. of Radiological Technology Shin Gu College) ;
  • Huh, Joon (Institute of Radiological Technology Listem Co., LTD)
  • Published : 2002.12.31

Abstract

We came to the following conclusion as the results of experiment on the image quality of mammography and the average glandular dose using 4 apparatuses at 3 hospitals in Seoul. 1. Whereas the measurement of half value layer showed no differences among the apparatuses, the measurement by an attenuation curve method showed some differences by 5.9%. There were 9.1% differences in the measurement by aluminum conversion method. 2. The basic density of an automatic exposure control unit must be D = 1.40, but there was no automatic exposure unit adjusted precisely at any hospitals. The unit at the B hospital exceeded the allowable limit by ${\pm}0.15$. 3. In the photographing using an automatic exposure control unit and the management of an automatic film processor using a sensitometer, most automatic film processors were well kept. But in some cases the mean value of a fluctuation coefficient exceeded the allowable limit. There is a need for more cautious management. 4. The image quality of breast phantom photography was affected by the screen/film system among the hospitals. 5. The average glandular dose at a breast of 4.2 cm thickness depended on the tube voltage, In the case of Mo/Mo, it was measured $0.26{\sim}1.39\;mGy$ less than ACR standard 3.0 mGy.

유방촬영에 대한 화질과 평균유선조직선량에 대하여 실험한 결과 반가층은 장치마다 큰 차이를 나타내지 않았으며, 자동노출기구를 사용할 경우 정확성에 문제가 있으므로 확인이 필요하다. 유방촬영에 사용되는 자동현상기 또한 꼭 관리가 필요하며, 유방촬영 시 사용되는 증감지/필름에 따라서 팬톰 촬영 식별능에 많은 영향을 받고 있으므로 선택 시 신중을 기해야 한다. 유방두께 4.2 cm에 대한 평균유선조직 선량은 관전압에 따라 달라지며, ACR기준인 3.0 mGy보다 훨씬 적은 $0.26{\sim}1.39\;mGy$로 나타났다. 이상에서 알 수 있듯이 유방촬영에서는 촬영장치의 성능뿐만 아니라 HVL, AEC, 자동현상기 관리, 증감지/필름 등 모두 관리가 되어야 하며, 이로 인해 평균유선조직선량은 더 줄일 수 있을 것이다.

Keywords