CORROSION PHENOMENA OF MAGNETIC ATTACHMENT USED IN ORAL ENVIRONMENT

구강 환경에서 사용된 자성 Dyna 어태치먼트의 부식현상

  • Kwak Jong-Ha (Department of Prosthodontics and Oral Biology Research Institute, College of Dentistry, Chosun University) ;
  • Chung Chae-Heon (Department of Prosthodontics and Oral Biology Research Institute, College of Dentistry, Chosun University) ;
  • Choe Han-Cheol (Department of Dental Laboratory Technology and Metallogical Engineering, Kwangyang University)
  • 곽종하 (조선대학교 치과대학 보철학교실 및 구강생물학연구소) ;
  • 정재헌 (조선대학교 치과대학 보철학교실 및 구강생물학연구소) ;
  • 최한철 (광양대학교 치과기공과 및 제철 금속학과)
  • Published : 2002.10.01

Abstract

The purpose of this study was to investigate further the mechanics of failure of magnets used for denture retention. Dyna magnets were retrived from denture that had failed after 34 months of clinical use. The magnets were observed and sectioned in order to analyse with high resolution scanning electron microscope. From this study, corrosion behaviors of used magnetic attachment were analysed. The results were as follows ; 1. In Nd-Fe-B based magnetic materials, the erosion-corrosion was started at ununiformed part of stainless steel cover. 2. Corrosion was initiated at weared stainless steel surface and then magnetic materials were spalled by corrosive solution. 3. Spatting was occurred in Nd-Fe-B magnet materials due to corrosion products and then corrosion rate was increased drastically. 4. Corrosion started from ununiformed stainless steel surface as well as welded zone. In conclusion, the failure of magnets may occur by either breakdown of the welding or breakdown of the encapsulating material. So, it is considered that the corrosion problem of dental magnetic materials could be solved to some extent with surface treatment of dental magnetic materials.

Keywords

References

  1. Hayes RJ, Gonten AS. Clinical application of rare-earth magnets. General Dentistry, 1990: 357-360
  2. Smith GA. Laird RE, Grant AA. Magnetic retention units for overdentures. J Oral Rehabil 1983:10:481-488
  3. Burns DR, Unger JW, Elswick RK, Beck DA. Prospective clinical evaluation of mandibular implant overdentures: Part 1-retention, stability, and tissue response. J Prosthet Dent 1995:73:354-363
  4. Carlyle LW, Duncan JM, Richardson JT, Garcia L. Magnetically retained implant denture. J Prosthet Dent 1986:56:583-586
  5. ackson TR. The application of rare earth magnetic retention to osseointegrated implants. Int J Oral Maxillofac Imp 1986: 1:81-92
  6. McCartney JW. Osseointegrated implant-supported and magnetically retained ear prosthesis: A clinical report. J Prosthet Dent 1991:66:6-9
  7. Sandler JP. An attractive solution to unerupted teeth. Am J Orthod Dentofac Orthop 1991: 100: 489-493
  8. Gianelly AA, Vaitas AS, Thomas WM. The use of magnets to move molars distally. Am J Orthod Dentofac Orthop 1989:96: 161-167
  9. Kalra V, Burstone CJ, Nanda R. Effects of a fixed magnetic appliance on the dentofacial complex. Am J Orthod Dentofac Orthop 1989:95:467-478
  10. Drago CJ. Tarnish and corrosion with the use of intraoral magnets. J Prosthet Dent 1991 :66:536-540
  11. Naert I, Quirynen M, Theuniers G, van Steenberghe D, Prosthetic aspects of osseointegrated fixtures supporting overdentures. A 4-year report. J Prosthet Dent 1991: 65:671-680
  12. Tsutsui H, Kinouchi Y, Sasaki H, Shiota M, shita T, Studies on the Sm-Co magnets as a dental material, J Dent Res 1979:58:1597-1606
  13. Vardimon AD. Graber TM, Drescher D. Bourauel C. Rare earth magnets and impaction, Am J Orthod Dentofac Orthop 1991:100:494-512
  14. Kitsugi A Okuno O, Nakano T, Hamanaka H, Kuroda T. The corrosion behaviour of Nd2Fe14B and SmCo5 magnets. Dent Mater J 1992:11:119-129
  15. Ko YM, Choe HC, Chung CH, Jung HS. Choi SW. A study on the development of safety for dental magnetic attachment materials. J Korean Rec Soc Dent Mater 1999:26(4):337-357
  16. Reley MA, Williams AJ. Speight JD, Walmsley AD, Harris IR. Investigations into the failure of dental magnets . Int Journal Prosth. 1999: 12:249-253
  17. Angelini E. Pezzoli M. Zucchi F. Corrosion under static and dynamic conditions of alloys used for magnetic retention in dentistry. J Prosthet Dent 1991 :65:848-853
  18. Sohn BS, Chang IT, Heo SJ. Keak JY, The electrochemical study on corrosion resistance of various dental magnetic attachments. J Korean Academy Prosthodont 2001 :39(4): 336-350
  19. Lim YS. Kim YS. A study on the physical properties and biological characteristics of dental magnetic attachments. Seoul National University, Graduate School Thesis 1999: 1-29
  20. Shreir LL. Corrosion vol. 1. 'Metal/Environmental reactions', 6:11, Newness Butterworths[London], 1972.
  21. Donald P : 'Hand book of stainless steel', 4:1, MacGraw-Hill[New York], 1977.
  22. 李鶴烈 : 金屬腐蝕工學, 淵鏡文化史, 105, 1996
  23. Nakano T. Okuno O, Hamanaka H. Released elements from Sm-Co magnet and dental casting ferromagnetic alloy by corrosion. Reports of the Institute for Medical and Dental Engineering 1988: 22: 17-22 (In Japanese)