DOI QR코드

DOI QR Code

Development of Rain Gauge and Observation Error

우량계 개발과 측정 오차

  • 김대원 (대구가톨릭대학교 환경과학과) ;
  • 이부용 (대구가톨릭대학교 환경과학과)
  • Published : 2002.10.01

Abstract

A new method of automatic recording raingauge is developed to measure rainfall 1200mm full scale with high accuracy and resolution. The principle of new instrument is to detect a weight change of a buoyant weight according to a change in water level of raingauge measured by the use of a strain gauge load cell. This method has the advantage of increasing measurement accuracy, since no moving equipment is used. Laboratory test of the instrument was recorded 0.4% error of 190mm rainfall amount. The validity of new instrument was examined by comparing its measured values with values recorded by automatic weather station on June 24 to 25 2001 at Daegu Meteorological Station, when there is 148.3mm rainfall amount. In spite of much rainfall there is only 0.77mm difference of total rainfall amount. This instrument was accomplished high accuracy and resolution at field test in much rainy day.

Keywords

References

  1. 건설교통부, 2000, 한국하천일람, 12pp.
  2. 한국수자원 공사 홈페이지, http://www.kowaco.or.kr
  3. Wada, 1911, Korean Meteorological Observatory Quart, J. Roy. Meteorol. Soc., 37, 83-86.
  4. 이부용, 1999, 측정 분해능이 0.1mm인 우량계의 개발에 관한 연구, 한국환경과학회지, 8(4), 419-422.
  5. 한국전자통신연구원, 1998, 위성통신 강우환경 분석기술 연구, 한국전자통신연구원 연차보고서, 120pp.
  6. M. Leroy, 2000, Calibration and Control Methods for Tipping-Bucket raingauges, Instruments and Observing Methods Report No. 74, 92-95.
  7. 이부용, 1998, 증발 기록계 개발에 관한 연구, 대구효성가톨릭대학교 연구논집, 57, 239-243.
  8. 이부용, 박병윤, 1997, 0.1mm급 우량계 개발에 관한 연구(LOAD CELL 특성에 관하여), 대구가톨릭대학교 응용과학연구소논문집, 5, 93-97.