Source Current Reconstruction Based on MCG Signal

심자도 신호를 이용한 전류원 재구성

  • Published : 2002.01.01

Abstract

When applying a SQUID system for diagnosing heart disease, it is informative to obtain the source current distributions from the measured MCG (magnetocardiogram) signals since the bioelectric activity in the heart is generally represented by distributed current sources. In order to estimate the Primary current distribution in a heart, the minimum norm estimate was computed, assuming a source plane below the chest surface. In the simulation, current distributions, which were computed for the test dipoles represented well the essential feature of the test-current configurations. Source current reconstruction was performed for MCG signal of a healthy volunteer, which was recorded using a 40-channel SQUID system in a magnetically shielded room. It was found that the obtained current distribution is consistent with the electrical activity in a heart.

Keywords