A 3-D Measuring System of Thermoluminescence Spectra and Thermoluminescence of CaSO4 : Dy, P
열자극발광 스펙트럼의 3차원 측정 장치와 CaSO4 : Dy, P의 열자극발광
- Lee, Jung-Il (Department of Physics, Kyeong-Sang University) ;
- Moon, Jung-Hak (Department of Ophthalmic Optics, Masan College) ;
- Kim, Douk-Hoon (Department of Ophthalmic Optics, Masan College)
- Published : 2001.12.31
Abstract
In this paper, a three-dimensional measuring system of thermoluminescence(TL) spectra based on temperature, wavelength and luminescence intensity was introduced. The system was composed of a spectrometer, temperature control unit for thermal stimulation, photon detector and personal computer for control the entire system. Temperature control was achieved by using feedback to ensure a linear-rise in the sample temperature. Digital multimeter(KEITHLEY 195A) measures the electromotive force of Copper-Constantan thermocouple and then transmits the data to the computer through GPIB card. The computer converts this signal to temperature using electromotive force-temperature table in program, and then control the power supply through the D/A converter. The spectrometer(SPEX 1681) is controlled by CD-2A, which is controlled by the computer through RS-232 communication port. For measuring the luminescence intensity during the heating run, the electrometer(KEITHLEY 617) measures the anode current of photomultiplier tube(HAMAMATSU R928) and transmits the data to computer through the A/D converter. And, we measured and analyzed thermoluminescence of
본 논문에서는 열자극발광 스팩트럼을 온도, 파장, 발광강도의 3차원으로 측정할 수 있는 장치를 소개하고, 고감도 TLD인