Physical Therapy Korea (한국전문물리치료학회지)
- Volume 8 Issue 3
- /
- Pages.35-41
- /
- 2001
- /
- 1225-8962(pISSN)
- /
- 2287-982X(eISSN)
Reliability of Plantar Pressure Measures Using the Parotec System
Parotec System을 이용한 족저압 측정의 신뢰도
- Roh, Jung-Suk (Dept. of Physical Therapy, Hanseo University) ;
- Kim, Tack-Hoon (Dept. of Physical Therapy, Hanseo University)
- Published : 2001.09.17
Abstract
In-shoe measurement systems allow the clinician and researcher to examine the pressure parameters within the shoe. The purpose of this study was to investigate the test-retest reliability of plantar pressures using the Parotec system over speeds and plantar regions. Seventeen healthy subjects were recruited for the study. Sampling rate was 100 Hz, and data of six variables (pressure on medial heel, lateral heel, 1st metatarsal head, 5th metatarsal head, and great toe and total impulse) were collected in four different gait speed (1.0 m/sec, 1.5 m/sec, 2.0 m/sec, and comfortable walking speed) in each day. The result indicates fair to excellent reliability between the two day test. Intraclass correlation coefficients (ICCs) ranged from .693 to .979, and range of reliability was similar depending on the speed and plantar region. In most cases, data recorded by the Parotec systems provide good evidence for the reliability.