The Validity Analysis between Measurement Method of Subglottic Air Pressure

성문하압 측정방법의 타당도 분석

  • 박상희 (대구대학교 재활과학과 언어치료) ;
  • 정옥란 (대구대학교 언어치료학과) ;
  • 석동일 (대구대학교 언어치료학과)
  • Published : 2001.09.01

Abstract

The purpose of the study was to examine a method most pertinent to measure subglottic air pressure. Subglottic air pressure and loudness analyses were performed on vowels /a/, /i/ and consonant /p/ in 12 normal subjects using. Aerophone II voice function. The experimental contexts were, therefore, /i:pi:pi:/ and /a:pa:pa:/. The subjects produced the intervocalic /p/ in 4 different situations: 1) /i:pi:pi:/ with voiceless /p/, 2) /i:pi:pi:/ with voiced /p/, 3) /a:pa:pa:/ with voiceless /p/, and 4) /a:pa:pa:/ with voiced /p/. A t-test and a correlation analysis revealed the following results. First, when we measured subglottic air pressure by /i:pi:pi:/, voiceless /p/ was significantly different from voiced /p/. Second, when we measured subglottic air pressure by /a:pa:pa:/, voiceless /p/ was significantly different from voiced /p/. Therefore, it was concluded that voiceless /p/ produced more accurate subglottic air pressure and clinicians needed to have patients produce accurate /p/ when measuring subglottic air pressure using Aerophone II.

Keywords