Abstract
Comparison of sulfur dioxide primary standard gases of the Korea Research Institute of Standards and Science (KRISS, Korea) and the National Physical Laboratory (NPL, UK) was performed. 100 ${\times}$10-6 mol/mol and 1,000 ${\times}$10-6 mol /mol primary standard gases (designated NPL S115 and S114, respectively) prepared gravimetrically and validated in NPL were used as transfer standards. Transfer standards were analyzed by NDIR sulfur dioxide analyzer and compared with KRISS PSM 112-03-624 and PSM 112-03-625 prepared gravimetrically. Adsorption corrected relative deviations of the primary standard gases were agreed to within 0.1%, and this agreement is within the expanded uncertainties (k = 2) of the primary standards at the two laboratories.