INTRAPULPAL TEMPERATURE CHANGE OF GLASS IONOMER ACCORDING TO LIGHT CURING INTENSITY AND CURING TIME

광중합기의 광도와 시간에 따른 글래스 아이오노머의 치수내 온도변화

  • 김희량 (전북대학교 치과대학 치과보존학교실 및 구강생체과학연구소) ;
  • 이형일 (전북대학교 치과대학 치과보존학교실 및 구강생체과학연구소) ;
  • 이광원 (전북대학교 치과대학 치과보존학교실 및 구강생체과학연구소) ;
  • 이세준 (전북대학교 치과대학 치과보존학교실 및 구강생체과학연구소)
  • Published : 2001.09.01

Abstract

When cavity floor is near the pulp, polymerization of light-activated restorations results in temperature increase. This temperature increase cause by both the exothermic reaction process and the energy absorbed during irradiation. Therefore instating base is required. Most frequently used insulating base is glass ionmer. The purpose of this study was to evaluate intrapulpal temperature changes of glass ionomer according to various curing intensity and curing time. Caries and restoration-free mandibular molars extracted within three months were prepared Class I cavity of 3$\times$6mm with high speed handpiece. 1mm depth of dentin was evaluated with micrometer in mesial and distal pulp horns. Pulp chambers were filled with 37.0$\pm$0.1$^{\circ}C$ water to CEJ. Chromium-alumina thermocouple was placed in pulp horn for evaluating of temperature changes. glass ionomer material was placed in 2mm. total curing time was 40s: continuous 40s, intermittent 20s, intermittent 10s. Glass ionomer material was cured with 300mW/$\textrm{cm}^2$, 550mW/$\textrm{cm}^2$ light curing unit. The results were as follows : 1. Temperature in pulp increased as curing unit power is increased. 2. Temperature in pulp more increased continuous emission than intermittent emission.