MR Imaging of Slow-flow Using a Flow Phantom

유동모형을 이용한 저속유동의 자기공명영상

  • Dae-Cheol Cheong (Departments of Radiology, College of Medicine, Catholic University of Taegu) ;
  • Kyung-Jae Jung (Departments of Radiology, College of Medicine, Catholic University of Taegu) ;
  • Young-Hwan Lee (Departments of Radiology, College of Medicine, Catholic University of Taegu) ;
  • Nak-Kwan Sung (Departments of Radiology, College of Medicine, Catholic University of Taegu) ;
  • Duck-Soo Chung (Departments of Radiology, College of Medicine, Catholic University of Taegu) ;
  • Ok-Dong Kim (Departments of Radiology, College of Medicine, Catholic University of Taegu) ;
  • Jong-Ki Kim (Departments of Radiology, College of Medicine, Catholic University of Taegu)
  • Published : 2001.12.01

Abstract

Purpose : To find sensitivity of MRI imaging methods to slow flow phantom study was performed with conventional Spin-Echo, gradient echo based Phase Contrast, fast GRASS, and heavily T2-weighted Fast Spin Echo pulse sequences. Materials and Methods : A siphon driven flow phantom was constructed with a ventriculo-peritoneal shunt catheter and a GE phantom to achieve continuous variable flow. Four different pulse sequences including Spin-Echo, Phase Contrast, GRASS and Heavily T2-weighted Fast Spin Echo were evaluated to depict slow flow in the range from 0.08 ml/min to 1.7 ml/min and to compare signal intensities between static fluid and flowing fluid. Results : In the slow flow above 0.17 ml/min conventional Spin-Echo showed superior apparent contrast between static and flowing fluid while GRASS was more sensitive to the very slow flow below 0.17 ml/mim. It was not accurate to calculate flow and velocity below 0.1 ml/min with a modified PC imaging. Conclusion : Four different MR pulse sequences demonstrated different sensitivity to the range of slow flow from 0.08 ml/min to 1.7 ml/min. This finding may be clinically useful to measure CSF shunt flow or detecting CSF collection and thrombosis.

목적 : 고식적 스핀에코, 위상 대조 경사에코, 고속 GRASS, 중T-2강조 고속스핀에코 연쇄를 각각 유동모형에 적용시켜, 저속유동에 대한 자기공명영상의 민감도를 찾아내고자 한다. 대상 및 방법 뇌실-복강 단락 도관과 GE 모형으로 이루어진 유동모형으로 싸이폰 효과에 의한 지속적인 가변 유속의 흐름을 내보내고, 각각의 유속에 위의 네 가지 자기공명영상연쇄를 적용시켰다. 0.08 ml/min 에서 1.7 ml/min 범위의 유량의 흐르는 액체와 정지된 액체에서 획득된 자기공명영상의 신호강도를 비교한 결과를 요약하면 다음과 같다. 결과 0.17 ml/min 이상의 느린 흐름에서는 고식적 스핀에코영상이 가장 우수한 정지-유동 액체사이의 겉보기대조를 보이나, 그 이하의 매우 느린 흐름에서는 GRASS 영상이 더 높은 민감도를 보였다. 결론 : 4가지 자기공명펄스연쇄는 0.08 ml/min에서 1.7 ml/min 범위의 유량의 흐름에서 서로 다른 민감도를 보였다. 이 결과는 임상적으로, 단락수술후의 뇌척수액의 흐름이나 혈관내의 혈전에 의한 흐름의 변이 등의 인체내의 저속유동을 검출하는 비침습적방법으로 유용하게 적용될 수 있을 것이다.